CONFIDENTIAL
VERSION
2.0
34 of 38
Conducted RF
a)
IEC 61000-4-6
3Vrms,
0.15 MHz
–
80 MHz
and
6Vrms,
in ISM and amateur
radio bands between
0,15 MHz and 80
MHz
3Vrms,
0.15 MHz
–
80 MHz
and
6Vrms,
in ISM and amateur
radio bands between
0,15 MHz and 80
MHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of Nukute Collare,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.7 GHz,
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Radiated RF EM fields
IEC 61000-4-3
10V/m Home
healthcare at 80 MHz
–
2700 MHz, AM
Modulation
3V/m at
2.7 GHz
–
6 GHz, AM
Modulation
10V/m Home
healthcare at 80 MHz
–
2700 MHz, AM
Modulation
3V/m at
2.7 GHz
–
6 GHz, AM
Modulation
a) The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13
13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are:
1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz to 7.3 MHz; 10.1 MHz to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07
MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz and 50 MHz to 54 MHz.
Содержание Nukute Collare