Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
–
guidance
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power supply lines
1 kV for input/output lines
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the THixyz series requires
continued operation during
power mains interruptions, it is
recommended that the
THixyz series be powered from
an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.