TRAINING UNIT

 

If you are using a training unit, please make sure to repack 
it according to the illustration.

 

 

 

 

 

 

 

 

SERVICE INTERVALLS

 

Once a year, or when used, NCCD shall be returned to NUI 
for maintenance. You will be contacted by NUI when it

time for service.

 

A new device will be sent to replace the unit that is

 

scheduled for maintenance.

 

 

RETURN AFTER USE

 

NCCD shall be returned to NUI after being used on a

 

patient. Please return complete unit in transport case to:

 

NUI AS

 

Gravdalsveien 245

 

5165 Laksevåg

 

Norway

 

Содержание NCCD

Страница 1: ...compression device to be used as a substitute to manual chest compression during CPR NCCD is powered by compressed gas and designed for use in hyperbaric environments NUI COMPACT CHEST COMPRESSION DEVICE NCCD USER MANUAL nui no January 2022 ...

Страница 2: ...djustable support strap 13 Extension hose 150 cm 14 Scissors for cutting fabric NCCD COMPONENTS 1 2 3 4 6 5 7 8 9 10 11 10 13 Parts 1 3 Compression unit 1 Piston house 2 Bottom plate 3 Piston foot 4 Hand control unit 5 Compression release handle 6 Pressure hoses 100 cm 7 Pressure hose 200 cm 12 14 ...

Страница 3: ...4 Do not compress the hand control unit 4 5 while connecting gas The NCCD piston foot 3 can move and cause a pinching hazard Apply NCCD before gas is connected Incorrect positioning will reduce quality of compressions and may cause injury to the patient MALFUNCTION If proper application and use of NCCD is not possible start manual CPR immediately Recommended checks and actions when possible 1 Veri...

Страница 4: ...e Hanson 600 series female coupling If your gas connection is of a different model please contact NUI Do not modify the NCCD in any way Do not use NCCD if the case seal is broken or missing Does not apply to the training units The NCCD will function while submerged in water INSTRUCTIONS For further instructions images and videos please visit nui no nccd If proper application and use of NCCD is not...

Страница 5: ...ssure release handle 5 is not compressed NCCD is to be used with pressurised HeliOx as drive gas at depth Compressed air can be used at surface Use of air as drive gas can significantly limit the NCCD functionality in a hyperbaric environment NCCD can be used along with defibrillator pads from an AED Make sure to place the pads under the NCCD chest belt After use After use make sure to disconnect ...

Страница 6: ...6 A Undress the patient s upper body Cut clothes using included scissors 14 if needed B Apply NCCD as instructed on the following pages Image show NCCD applied APPLICATION OF NCCD ...

Страница 7: ...EN NCCD CASE Break sealing It requires 6 kg of force 2 TAKE OUT NCCD Remove the NCCD with chest belt 9 black from the case The belt is pre attached to the right side of the bottom plate 2 APPLICATION OF NCCD ...

Страница 8: ...NT S CHEST Place the NCCD compression unit 1 3 on patient chest so that the piston foot 3 is located where your hands would have been placed during manual CPR Make sure not to place the piston foot to far down towards the tip of the sternum or too high up on the chest This could cause injury to the patient or and cause insufficient compressions APPLICATION OF NCCD 3 4 ...

Страница 9: ...ure it s firmly fastened on both sides Use red markings 11 to center the belt c Adjust both sides if necessary Confirm that the NCCD is firmly applied to the patient s chest The NCCD shall not compress the chest by application Confirm correct placement of the piston foot 3 according to the hand placement used when you are preforming manual CPR APPLICATION OF NCCD a c b ...

Страница 10: ...belt a Remove the adjustable support strap 2 from the package and hang it around the patient neck b Attach the carabiner hooks to the slots in the NCCD plate 2 as shown in image b and c The strap length and tension needs to be adjusted so that it supports the NCCD not causing an angle perpendicular to patient frontal plane or causing a rotation of the NCCD c Verify that the adjustable strap is att...

Страница 11: ...rm 10 bar over ambient gas drive pressure Confirm that gas supply has approx the same ppO2 as compartment atmosphere 7 CONNECT TO GAS Locate quick connector QC 8 for gas connection Insert male QC on NCCD into gas outlet female QC as you would a BIBS mask USE OF NCCD ...

Страница 12: ...t piston is compressing the chest according to CPR requirements Repeat and continue preforming chest compressions in accordance with current CPR protocol Not fully compressing and releasing the handle on the control unit will cause insuficcient chest compressions Strive to complete full compressions og decompressions using as much time to compress the handle as to release the handle on the control...

Страница 13: ...ith a cloth with detergent and warm water until visibly clean Then use sufficient disinfection like alcohol 75 or Virkon NCCD chest belt 9 The belt can be machine washed at 90 C 194 F up to one hour program Shirking of the fabric when washed 1 If the belt is contaminated with biological matter it should be cleaned with soap and water or machine washed as described above Then cleaned with a suffici...

Страница 14: ... NCCD shall be returned to NUI for maintenance You will be contacted by NUI when it s time for service A new device will be sent to replace the unit that is scheduled for maintenance RETURN AFTER USE NCCD shall be returned to NUI after being used on a patient Please return complete unit in transport case to NUI AS Gravdalsveien 245 5165 Laksevåg Norway ...

Страница 15: ...ating temperature 5 C 41F to 60 C 140F Sound level at operating distance 80 dB Storage temperature 5 C 41F to 40 C 104F Maximum ambient pressure 30 bar 435 psi 300 msw Requirements Gas drive pressure 10 bar 145 psi over ambient Chest circumference 90 130 cm 31 51 in Dimensions Chest belt 10 L150 cm W11 cm NCCD 1 11 weight 1950 g Case size W265 x D240 x H174 mm Case empty weight 1200 g Case with NC...

Страница 16: ...es the subsea industry with a great diversity of products and services ranging from research development and advisory to hyperbaric contingency engineering pressure and equipment testing Our aim is to continuously contribute to improving equipment safety lowering health risks and raising awareness around hyperbaric operations www nui no post nui no 47 55 94 28 00 N C C D ...