Chapter 4
ESPrit 3G User Guide
♦
37
Warnings and Precautions
This section describes the warnings and precautions that
apply to your cochlear implant system. Read this section
carefully to ensure that you understand the care of your
system.
Discuss these warnings and precautions with your
physician before undergoing any major medical
procedure.
Warnings
Medical Treatments Generating Induced
Currents
Some medical treatments generate induced currents that
may cause tissue damage or permanent damage to the
cochlear implant. Warnings for specific treatments are
given below.
· Electrosurgery:
Electrosurgical instruments are capable
of inducing radio frequency currents that could flow
through the electrode array.
Monopolar
electrosurgical
instruments must
not
be used on the head or neck of a
cochlear implant patient as induced currents could cause
damage to cochlear tissues or permanent damage to the
implant.
Bipolar
electrosurgical instruments may be
used on the head and neck of patients, however, the
cautery electrodes must not contact the implant and
should be kept more than 1 cm (~ ½ in.) from the
extracochlear electrodes.
N94722F ISS2 body d02 MAS.p65
1/7/02, 3:23 PM
37
Содержание ESPrit 3G
Страница 1: ...User Guide for the ESPrit 3G Speech Processor and Accessories...
Страница 2: ...User Guide Part Number N94722F Issue 2 January 2002 for the ESPrit 3G speech processor and accessories...
Страница 3: ......
Страница 7: ...Chapter 4 6 ESPrit 3G User Guide...
Страница 19: ...Chapter 4 18 ESPrit 3G User Guide...
Страница 25: ...Chapter 4 24 ESPrit 3G User Guide...
Страница 29: ...Chapter 4 28 ESPrit 3G User Guide...
Страница 45: ...Chapter 4 44 ESPrit 3G User Guide...
Страница 48: ...Cochlear Ltd ABN 96 002 618 073 14 Mars Road Lane Cove NSW 2066 Australia...