OM0211
Page 25 of 48
Rev 3 March 2013
8.6
Front Component Panel
The front component panel contains the following functions. See BCD-15180
8.6.1
CO2 Supply Filter
The CO2 supply filter is installed at the factory and should be changed following the instructions found in
Maintenance Section 10.
8.6.2 Air Inlet Filter - All models
The Incubator is provided with clear vinyl tubing and a 0.3-micron HEPA filter. This is a free air supply.
Note: DO NOT CONNECT TO A PRESSURIZED SOURCE.
8.6.3 Chamber sample port
The Sample port is provided for checking the CO2 level in the chamber to monitor the calibration the Sensor and
determine if it need adjustment.