OM0312
33
11.0 Electrical/Environmental Requirements
11.1 Electrical: (
Supply voltage fluctuations not to 10%)
NU-608-400E 230 VAC,
50/60 Hz,
1 Phase,
5 Amps
Optional integral 230 VAC, 50/60 Hz, 1 Phase, .9 Amps
auto lift base
*UL Listed or CE Certified
11.2 Operational Performance
(for indoor use only)
Environment Temperature Range:
60
F - 85
F (15
C - 30
C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50% relative humidity at 40°C
Environment Altitude:
6562 Feet (2000 meters) maximum
11.3 Light Exposure
Standard LED Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
11.4 Installation Category:
2.0
Installation category (overvoltage category) defines the level of transient overvoltage, which the instrument is
designed to withstand safely. It depends on the nature of the electricity supply and its overvoltage protection
means. For example, in CAT II, which is the category used for instruments in installations supplied from a supply
comparable to public means, such as hospital and research laboratories and most industrial laboratories, the
expected transient overvoltage is 2500 V for a 230 V supply and 1500 V for a 120 V supply.
11.5 Pollution Degree
: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution
degree 2 assumes that normally only non-conductive pollution such as dust occurs with the exception of
occasional conductivity caused by condensation.
11.6 Chemical Exposure
Chemical exposure should be limited to antibacterial surface disinfectants used for cleaning and disinfecting.
USE OF
CHLORINATED OR HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL.
Equipment decontamination
can be accomplished by non-condensing gas or vapor Paraformaldehyde, Hydrogen Peroxide or Chlorine Oxide following
NSF/ANSI 49,
Annex I-2 (formerly Annex G).
11.7 EMC Performance (classified for light industrial)
Emission:
EN61326
Immunity:
EN61326
Warning:
Class A equipment is intended for use in an industrial environment.
In the documentation for the user, a statement shall be included drawing
attention to the fact that there may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted as
well as radiated disturbances.
!