4.1.15 Menu 15
– Electronics zero
Disconnect the sensor and place a short at the sensor input to simulate 0%
oxygen.
•
Press the MENU button to open the menu system.
•
Using the NEXT and PREVIOUS buttons select menu option:
E:15
•
Press ENTER.
•
Press ENTER to zero the sensor, ‘
– – – –
‘will be displayed to
confirm the sensor zero has been performed.
Note: Pressing the MENU button rather than the ENTER
button exits the zero feature without performing the
calibration.
•
Press the MENU button to close the menu system.