4

ENGLISH  

GERMAN  FRENCH  SP

ANISH  IT

ALIAN

5

Insert

Handpiece

Motor

Tighten

Spanner
Wrench

Blade

U-cut

Chuck

Chuck Nut

Fig. 2

*Picture shows SGR2-E.

*Picture shows SGR2-E.

2. Connection of Handpiece and Motor

2-1 Connecting the Handpiece to the Micromotor

Simply push the handpiece directly onto the Micromotor. (Fig. 1)

3. Mount/Removal of Blade

Mount the NSK reciprocating / oscillating / sagittal Blades (option).

3-1 Mounting Blade

1) Loosen the Chuck Nut approximately 2-3 turns by hand.
2) 

SGR2-E, SGT2-E

Align the blade and the chuck slit then insert the blade until it stops.

SGO2-E

Insert the D-shaped shank into the chuck, align shapes of the chuck 
and shank by turning lightly, and insert until it stops.

3) Tighten the Chuck Nut with the Spanner Wrench. (Fig. 2)

CAUTION

Check the blade is secure before use.

CAUTION

Check the connection is secure by gently 
pulling the handpiece and Micromotor.

2-2 Removing Handpiece from Motor

Hold the handpiece and Micromotor, and pull 
the handpiece direct from the Micromotor.
(Fig. 1)

CAUTION

· Note that only E-type Micromotors according to ISO 3964 can be 

connected.

· Connect and/or disconnect the handpiece without a blade inserted.
· Do not connect and/or disconnect the handpiece until the 

Micromotor is completely stopped.

Fig. 1

Содержание SGR2-E

Страница 1: ...001 OM H0187E BONE SURGERY MICRO SAW HANDPIECE 10 00 00 N SGR2 E SGO2 E SGT2 E The EU directive 93 42 EEC was applied in the design and production of this medical device...

Страница 2: ...ct your authorised dealer Entering foreign materials into bearing may cause OVERHEATING Do not attempt to disassemble the handpiece or tamper with the mechanism Do not deliberately transform external...

Страница 3: ...on for 2 minutes after continuous use for 3 minutes approximately Never move the chuck nut during use as this can cause injury Be sure to replace the saw blade only after the Micromotor completely sto...

Страница 4: ...t then insert the blade until it stops SGO2 E Insert the D shaped shank into the chuck align shapes of the chuck and shank by turning lightly and insert until it stops 3 Tighten the Chuck Nut with the...

Страница 5: ...6 5mm 9 2mm 9 2mm 10 5mm 10 5mm 15 5mm 30mm 45mm Blade thickness 0 3mm Blade thickness 0 35mm Blade thickness 0 35mm Blade thickness 0 35mm Blade thickness 0 35mm 15 5mm 15 5mm 15 5mm 30mm 45mm 45mm 2...

Страница 6: ...the E type Spray Nozzle in the rear of the handpiece Spray the lubricant into the handpiece 2 3 times for 2 3 seconds each time until blood or foreign material is not visible in the discharging lubri...

Страница 7: ...after each patient Always place the handpiece in the center or upper shelf of the chamber as the local temperature at the bottom of chamber could rise beyond the set value Do not wipe nor immerse the...

Страница 8: ......

Отзывы: