OM-H0187E 

BONE SURGERY MICRO SAW HANDPIECE

’10.00.00 N 

SGR2-E/SGO2-E/SGT2-E 

The EU directive 93/42/EEC was applied in the design and production of this medical device.

Содержание SGR2-E

Страница 1: ...001 OM H0187E BONE SURGERY MICRO SAW HANDPIECE 10 00 00 N SGR2 E SGO2 E SGT2 E The EU directive 93 42 EEC was applied in the design and production of this medical device...

Страница 2: ...ct your authorised dealer Entering foreign materials into bearing may cause OVERHEATING Do not attempt to disassemble the handpiece or tamper with the mechanism Do not deliberately transform external...

Страница 3: ...on for 2 minutes after continuous use for 3 minutes approximately Never move the chuck nut during use as this can cause injury Be sure to replace the saw blade only after the Micromotor completely sto...

Страница 4: ...t then insert the blade until it stops SGO2 E Insert the D shaped shank into the chuck align shapes of the chuck and shank by turning lightly and insert until it stops 3 Tighten the Chuck Nut with the...

Страница 5: ...6 5mm 9 2mm 9 2mm 10 5mm 10 5mm 15 5mm 30mm 45mm Blade thickness 0 3mm Blade thickness 0 35mm Blade thickness 0 35mm Blade thickness 0 35mm Blade thickness 0 35mm 15 5mm 15 5mm 15 5mm 30mm 45mm 45mm 2...

Страница 6: ...the E type Spray Nozzle in the rear of the handpiece Spray the lubricant into the handpiece 2 3 times for 2 3 seconds each time until blood or foreign material is not visible in the discharging lubri...

Страница 7: ...after each patient Always place the handpiece in the center or upper shelf of the chamber as the local temperature at the bottom of chamber could rise beyond the set value Do not wipe nor immerse the...