· When operating this motor always consider the safety of the patient.
· The motor is designed only for clinical dental use by qualified personnel.
· This motor can not operate without exclusive control unit.
· Do not allow any impact on to the motor. Do not drop it. It is cause a malfunction.
· Check for vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop
using the motor immediately and contact authorized dealer.
· Do not exceed the drive motor speed recommended the handpiece (Option) manufacturers.
· Prior to use, always check the handpiece tightness, vibration, noise and overheatin. If any abnormalities are detected, stop
using immediately and contact authorized dealer.
·
Do not use contaminated air (by dust or moisture). If the air contains water or dust, it might cause a malfunction or
overheat.
· Do not use this motor under strong stress for long time. It is cause overheat.
· Do not connect or disconnect the cord until the drive motor has completely stopped.
· Do not connect / disconnect the handpiece during operation.
· Verify that the Speed Control Switch is adjusted within the allowable speed before use.
· Air Requirements: dry, free from contamination and oil. Use a compressor with a dry air system. Install an air filter if
necessary. Blow out the lines before installation.
· Do not Autoclave (or any other high temperature Sterilization) Motor Cord.
· The user shall be responsible for operation and maintenance.
· The operator is responsible for correct operational control, maintenance and inspection.
· Surface temperature shall be more 60°C without using a coolant air, to avoid this event, be sure to use coolant air.