Maintenance 

 

 

Maintenance  is 

not

  required  except  that

the  Nuclestat®  must  be  kept  clean  by
blowing dry, filtered, compressed air over
its surface. 

 

 

Calibration is 

not

 required. 

 

 

To  ensure  maximum  static  removal
efficiency,  the  Nuclestat®  must  be 
replaced  annually  regardless  of  use.    The
replacement date is one year after the date
manufactured 

(printed 

on 

each

Nuclestat®).  NRD will remind customers 
of the replacement date. 

 
 

Trouble Shooting Guide 

 

Symptom 

Probable Cause 

Corrective Action 

 

1.

 

Static charges are not 
eliminated 

 
 
 
 
 
 
 
2.

 

Static charges are 
eliminated only on one 
side of the material being 
processed 

 

Nuclestat

® 

is out of date and 

needs replacement 
 
Dirt buildup on the 

Nuclestat® 

 
 
Nuclestat

®

 or equipment is not 

properly grounded 
 
Material is in contact with 
equipment or additional 
material layers

 

 

Call NRD for a replacement 
Nuclestat

®

 

 
Remove dirt using clean, dry, 
compressed air 
 
Secure

 

Nuclestat

® 

or 

equipment to ground potential 
 
Freely suspend material being 
processed 

 

 
 

For additional assistance call NRD Customer Service toll-free at 

 

1-800-525-8076

 

 
 

 