INST6210 – 4/09 

Installation / Operation / Maintenance 

Sentinel Benchtop Ionizing Blower 

Model 6210 

Changing Voltage

 

 

The input power voltage should be set to your specification 
as ordered and prior to shipping.  It is marked on the box 
label and the specification label affixed to the bottom of the 
Ionizer.  The voltage can be reset to a different AC voltage 
by removing the back of the case.  NOTE:  The AC power 
cord MUST always be disconnected before the Ionizer is 
disassembled.  Reconfigure the voltage jumpers as shown in 
the diagram(s) below to match your AC line voltage supply. 

 

Input voltage is selected with the two internal jumpers as 
shown below: 
 

115 volt Jumper Settings 

 

 
 

230 volt Jumper Settings 

 

 

Specifications 

 

Air Velocity:  

Two speed fan (250 lfpm with fan on high 

                         speed). 

 

Balance:  

±

 3 volt typical 

                 (Temperature range: 65

°

F to 80

°

F, 

                 Relative humidity: 15% to 65%) 

 

Ion Emission:  

Steady state DC with sensor feedback. 

 

Emitter Pin Electrodes:  

.050” diameter 316 stainless steel; 

                                           8 required. 

 

High Voltage Power Supply:  

5.5 kV DC nominal; 

                           (1) positive HVDC and (1) negative HVDC 

 

Input Power:  

AC line power, internally selectable for 

                         100/120 VAC, 50/60 Hz, 1 phase or 
                         200/240 VAC, 50/60 Hz, 1 phase 

 

Fuse:  

250 mA slow blow; located at AC inlet inside 

            rear cover; 2 required 

 

Chassis:  

Powder coated aluminum or steel housing 

 

Weight: 

 7.0 lbs. With cord; 7.4 lbs. Shipping weight with 

                Packaging 

 

Dimensions:  

Width:  6.76” (8.08” with knobs) 

                       Height:  9.82” (with stand) 
                       Deep:  3.77” 

 

Mounting:  

Bench top tilt adjust frame with locking knobs; 

                    can also be wall or shelf mounted. 

 

Ozone:  

Less than 0.005 ppm 

 

              

 

 

 

Power 
Jack 

 Voltage 
 Selection 
 Header 

 Jumpers 

Power 
Jack 

 Jumper 

 Voltage 
 Selection 
 Header 

Содержание Sentinel 6210

Страница 1: ...on and maintenance instruc tions sheet Installation Locate the Ionizer 12 to 48 from the target to be neutral ized for optimal performance making sure that the airflow will not be restricted The Ionizer comes with a stand so that it can be positioned on a flat surface aimed at the target and locked via adjustment knobs It can also be wall or shelf mounted to keep valuable bench space free for othe...

Страница 2: ... timum neutralization efficiency and operation cleaning should be performed on a regular basis once a month is rec ommended In addition to improving decay time cleaning also keeps the Ionizer balanced without the need for adjust ment NOTE The AC power cord MUST be disconnected before the Ionizer can be disassembled for maintenance First turn the Ionizer OFF and unplug the power cord Remove the bac...

Страница 3: ...or a Return Authorization number and ship ping address IMPORTANT NOTE In some cases if the balance sensing grill inside the blower becomes contaminated the Ionizer may not be able to detect an out of balance condition and may continue to run in a green light status The presence of contamination on the backside the side nearest the fan of the balance sensing grill can cause the Ionizer to overcompe...

Страница 4: ... Air Velocity Two speed fan 250 lfpm with fan on high speed Balance 3 volt typical Temperature range 65 F to 80 F Relative humidity 15 to 65 Ion Emission Steady state DC with sensor feedback Emitter Pin Electrodes 050 diameter 316 stainless steel 8 required High Voltage Power Supply 5 5 kV DC nominal 1 positive HVDC and 1 negative HVDC Input Power AC line power internally selectable for 100 120 VA...

Отзывы: