background image

R/I-T1DX User Manual 

_______________________________________________________________________________________________________________

 

___________________________________________________________________________ 
version 1.4 

 

 

page 7 

2.2. 

Frequency and Amplitude of Current Injection Pulse 

 
Frequency and amplitude of current injection are dependent on each other according to 

dt

dU

C

I

C

 

where I

c

 is the capacitive current, C is the capacitor in the amplifier headstage (3.3 pF) , dU/dt 

is the change of the voltage amplitude of the triangle over time. 
 
Example for frequency with fixed amplitude of the triangle (±10 V i.e. 20 V amplitude) 

dt

dU

C

I

C

 

or 

C

I

dU

C

dt

 

 

s

nA

V

F

dt

660

100

20

10

3

.

3

12





 

 
because of having a biphasic pulses this has to be multiplied by 2 to get the frequency 
 
660 µs *2 = 1.32 ms 

 758 Hz 

 
 

3. Technical Data 

 
R/I-T1DX 
 
Resistance measurement:  0.01…10 M

, display: XX.XX M

 

Current injection: 

0…±280 nA, symmetrical around baseline; display: XXX nA 

Output scaling: 

1 V / M

 or 1 V / 10 nA 

Output: resistance: 

50 

, range ±10 V 

Содержание R/I-T1DX

Страница 1: ...TION FOR THE R I T1DX ELECTRODE RESISTANCE TEST CURRENT INJECTION VERSION 1 4 npi 2013 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support npie...

Страница 2: ...__________________ ___________________________________________________________________________ version 1 4 page 2 Table of Contents 1 Safety Regulations 3 2 R I T1DX Resistance Test Current Injection...

Страница 3: ...is to be operated only at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a prot...

Страница 4: ...sulting in a rectangular current signal at the electrode For electrode resistance test the voltage drop at the electrode caused by the current signal is measured and the electrode resistance is calcul...

Страница 5: ...by 5 5 RESISTANCE CURRENT display Digital display showing the electrode resistance XX XX M range 0 01 M 10 M or the amplitude of the injected current XXX nA range 0 100 nA 6 Current nA resistance M LE...

Страница 6: ...Customized auxiliary connector 3 TTL IN connector BNC connector for remote control of operation with a TTL signal As long as the signal is HI 2 5V operation electrode resistance test or current inject...

Страница 7: ...itive current C is the capacitor in the amplifier headstage 3 3 pF dU dt is the change of the voltage amplitude of the triangle over time Example for frequency with fixed amplitude of the triangle 10...

Отзывы: