background image

EXT-02F User Manual 

___________________________________________________________________________ 
version 4.2 

 

 

page 10 

3.  Operation 

 
The  EXT-02F  can  be  operated  either  using  the  inputs  of  the  front  panel  (#16,  Figure  1)  or 
using a headstage. 
 
Very Important: You must not use inputs at the front panel and a headstage for one channel 
simultaneously!! If the BNC connectors are used the headstage must be disconnected. 
 
 
Also Important: If you encounter noise when operating the EXT-02F try to place the amplifier 
as  far  away  as  possible  from  the  power  supply!!  This  will  avoid  magnetic  coupling  of  the 
power supply to the amplifier! 
 
Extracellular  measurements  are  mostly  done  in  slices  or  in  vivo,  where  distortions  of  the 
signal  caused  by  other  instruments  and  the  animal  itself  are  very  common.  Additionally, 
extracellular signals are very small and have to be amplified enormously. The drawback is that 
noise  is  amplified  as  well.  Therefore,  each  channel  of  the  EXT-02F  is  equipped  with 
differential  input  that  minimizes  noise  pick-up.  Differential  means,  that  the  signal  for  the 
amplifier  is  the  difference  between  the  positive  (+)  (P

EL

  at  the  headstage)  and  negative  (-) 

(REF.  at  the  headstage)  input  of  the  amplifier.  This  results  in  canceling  of  all  signals  which 
both electrodes record, e.g. noise. 
 
For  differential  measurements,  both  inputs  are  connected  to  electrodes  using  cables  with 
grounded enclosure or electrode holders. + (P

EL

) is connected to the measuring electrode and – 

(REF.) to the reference electrode. If no headstage is used, the switches #14, Figure 1 have to 
be in the same position, i.e. both set to ON. The experimental chamber is grounded by an Ag-
AgCl pellet (or an AGAR bridge) connected to GND of the headstage (see Figure 4). 
 
Important:  If  differential  measurement  is  not  required  (single-ended  measurement 
configuration), the REF input must be connected to ground (GND at the headstage, see Figure 
4) or the –IN switch (#14, Figure 1) must be set to OFF. 
 
If no headstage is used, the switch corresponding to the unused INPUT (+ or -) has to be set to 
OFF (see #14, Figure 1). 
The  amplifier  is  in  an  undefined  state,  if  the  REF  is  left  open,  and  can  go  into  saturation 
making reliable measurements impossible. 

Содержание EXT-02F

Страница 1: ...E EXT 02F EXTRACELLULAR AMPLIFIER WITH FILTERS One or Two Channel Recordings VERSION 4 2 npi 2015 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 s...

Страница 2: ...______ version 4 2 page 2 Table of Contents 1 Safety Regulations 3 2 EXT 02F 4 2 1 Components 4 2 2 System Description 4 2 3 Description of the Front Panel 4 2 4 Description of the Rear Panel 8 2 5 He...

Страница 3: ...e check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Befor...

Страница 4: ...of the two recording channels are identical The EXT 02F has a differential input with high input impedance to avoid noise The output voltage signal can be HIGH PASS or LOW PASS filtered and is availa...

Страница 5: ...osition switch for selecting the corner frequency of the single pole HIGHPASS filter with 6 dB octave In OFF position the HIGHPASS filter is disabled DC recording Corner frequencies Hz 3 30 100 300 40...

Страница 6: ...gnals greater than 1 V will be converted into a sound NOISE With an amplification of 10000 this correspondents to a signal of 100 V at the electrode 9 INPUT MODE switch Switch for selecting the INPUT...

Страница 7: ...he HIGH PASS FILTER to 400 Hz Set the GAIN range switch to 1k and adjust the baseline to zero using the trimpot Set the GAIN range switch to 10k and adjust the baseline again if necessary 14 INPUT swi...

Страница 8: ...are used the headstage must be disconnected Also very Important Setting both switches to OFF means that both inputs are grounded i e measurement is impossible This applies also if a headstage is used...

Страница 9: ...ON A switch Switch that enables upper position or disables lower position the NOTCH filter optional 2 5 Headstage Option Figure 3 headstage optional and electrode holder optional and of the EXT 02F He...

Страница 10: ...al input that minimizes noise pick up Differential means that the signal for the amplifier is the difference between the positive PEL at the headstage and negative REF at the headstage input of the am...

Страница 11: ...______________________________ version 4 2 page 11 Figure 4 headstage connections A differential measurement B single ended measurement Hint The REF connector of the headstage can also be set to GND b...

Страница 12: ...Electrical Activity of Individual Neurons In Situ Extra and Intracellular Recording in U Windhorst and H Johansson eds Modern Techniques in Neuroscience Research Springer Berlin New York Lambert F M...

Страница 13: ...400 500 600 800 1k 3k 10k with OFF position LOWPASS Filter single pole attenuation 6 dB octave corner frequencies Hz 30 100 300 500 700 900 1 3k 1 5k 1 7k 3k 10k 20k NOTCH Filter Europe 50 Hz USA 60...

Отзывы: