ATTENTION! 
Read this instruction manual carefully before switching on the gas analyzer! 
 

This  manual  explains  how  to  use  the  gas  analyzer  toxic  and  fuel  gases  М 02-01, 

consider of operating principles and technical characteristics for correct appliance 

 
1 Purpose 
 
1.1  Analyzer  M 02-01  are  designed  for  rapid  automatic  continuous  measuring 

methane  concentrations,  oxygen  concentration,  carbon  monoxide  concentration,  carbon 
dioxide  concentration,  the  hydrogen  sulfide  concentration  in  the  atmosphere  for  general-
purpose  object class V-1a, and in underground  mines, including  dangerous by  gas or  dust 
and sudden outburst (blast). 

1.2 Type of gas analyzer - portable, automatic, multi-channel, for individual use. 
1.3 Operating conditions of gas analyzers: 
- the range of ambient temperature from  – 30 to + 40 °C, briefly in the ranges from 

30-40 °C below zero and 40-50 ° C above zero; 

- relative humidity up to 100 % at + 35 °C; 
-  the  range  of  atmospheric  pressure  from  87.8  to  119.7  kPa  (from  660  to  900  mm 

Hg); 

- the speed of the gas air flow to 8 m/s; 
- vibration with frequency (5-35) Hz and amplitude not more than 0.35 mm. 
1.4 Gas analyzers provide: 
- continuous measurement and digital indication of the monitored component; 
-  indication  and  signaling  (sound  and  light  intermittent  signals)  of  exceeding  the 

established  threshold  values  for  the  concentration  of  methane  (or  methane-hydrogen 
mixture), carbon monoxide, carbon dioxide, hydrogen sulfide, oxygen; 

-  continuous  automatic  recording  of  information  about  the  concentration  of  the 

measured  component  and  the  parameters  of  the  gas  analyzer  operation  in  real  time  in  the 
archival memory (the function of the "black box"); 

-  fixation  of  the  results  of  measurements  of  the  concentration  of  the  monitored 

component at the command of the operator (immediately or with a delay), storing them in 
the memory of the gas analyzer; 

- sampling and indication of the fixed values from the gas analyzer memory; 
-  transmission  of  information  stored  in  the  "black  box"  mode  or  separately  by  the 

operator's command, via the infrared port to the personal computer; 

- indication of the current date and time; 
- indication of the ambient temperature; 
- indication of atmospheric pressure; 
- fault indication; 
- charge management, indication and signaling about the discharge of the battery; 

Содержание M 02-01

Страница 1: ...GAS ANALYZER TOXIC AND FUEL GASES 02 01 Operating Manual Short version...

Страница 2: ...2 Contents Page 1 Purpose 3 2 Technical specifications 4 3 Operation and description 5 4 Preparation for work 8 5 Operating procedure 10 6 Safety requirements 13...

Страница 3: ...n with frequency 5 35 Hz and amplitude not more than 0 35 mm 1 4 Gas analyzers provide continuous measurement and digital indication of the monitored component indication and signaling sound and light...

Страница 4: ...f carbon dioxide volume fraction from 0 to 5 range of indications of the volume fraction of carbon dioxide from 0 to 100 range of ambient air temperature readings from 50 to 50 C range of atmospheric...

Страница 5: ...episodic measurements is 4000 records At filling of the specified volume new records are brought in a place of the first It is possible to clear format the memory of occasional measurements in the tu...

Страница 6: ...enings for supplying the gas mixture to the sensors 6 made in the form of a grate consisting of inclined plates preventing direct moisture penetration onto the sensors A light conducting zone 7 is loc...

Страница 7: ...7 Figure 1...

Страница 8: ...as analyzer will pass to the testing phase automatic zero or warm up and a view window for M 02 01 with measurement channels CH4 O2 CO H2S will appear on the display The upper part of the window shows...

Страница 9: ...05 along the carbon monoxide measurement channel from 0 ppm to 2 ppm through the hydrogen sulfide measurement channel from 0 ppm to 1 ppm and the reading on the channel for measuring the volume fract...

Страница 10: ...t sound intermittent signal appears he display illumination turns on and a blinking icon of the reached first or second threshold level appears on the screen in the gas code line respectively for M 02...

Страница 11: ...ith the illumination of the display screen and a flashing contour of the battery icon on the display screen After the warning signals the gas analyzer can work for another 25 minutes at a positive amb...

Страница 12: ...croll buttons to move the active marker to TURN OFF 5 5 3 Press the enter button of the keyboard In this case the window with the working menu will be replace by the deactivation window 15 24 08 08 09...

Страница 13: ...out work with the gas analyzer observe the requirements of the regulatory documents on occupational safety applicable in the industries where this gas analyzer is used 6 2 WARNING During operation pre...

Отзывы: