6000counts Intelligent scanning digital multimeter
6
/
7
measurement in the second cycle, and select the corresponding function measurement according to the measurement requirements. When measuring
the capacitance, the size of the measured capacitance will automatically select different ranges, and the measured value will be displayed on the LCD.
The capacitance measurement position is 10nF/100nF / 1uF/10uF/100uF/1mF/10mF/60mF.
Note:
1
)
When measuring the capacitance with the 10nF file, the screen display value may have a residual reading, which is the distributed capacitance
of the test pen, which is an accurate reading, which can be subtracted after the measurement;
2) When the large capacitance file measures severe leakage or breakdown capacitance, some values will be displayed and unstable; when
measuring large capacitance, the reading takes several seconds to stabilize, which is normal when measuring large capacitance;
3
)
Please fully discharge the capacitor before testing the capacitor capacity, otherwise it will enter the voltage measurement mode.
4
)
Unit
:
1F=1000mF 1mF=1000uF 1uF =1000nF
1nF=1000pF
2-3-4. AC and DC current (DC/ACV A)
Maximum measured pressure drop: 600mV; Overload protection: 10A
The specific operations are as follows:
1. The boot is displayed as an automatic scan status "----".
2. Insert the black test lead into the
“
COM
”
jack and the red test lead into the
“
V/
Ω”
jack; the test leads are reliably in contact with the test point.
3. When inputting more than 20mA current between "COM" and "10A", the meter will display the large component on the LCD according to the size
of the AC and DC components.
Note
:
1) Before the meter is connected in series to the loop to be tested, the power in the loop should be turned off first;
2) The maximum input current is 10A (depending on the insertion position of the red meter pens), not more than 10 seconds, excessive current will
cause the circuit to heat, or even damage the instrument;
3) When the test leads are plugged into the current input port, do not connect the test leads to any circuit, which will damage the fuses and
Instruments;
4
)
After the completion of all the measurement operation, you should first turn off the power and then disconnect the test leads and the measured
circuit connection, Especially to high current measurement.
5)
It is forbidden to input more than 36V DC, 25V AC voltage between the current jack and the "COM" jack.
2-3-5
.
NCV measurement
The specific operation is as follows:
1. Press and hold the
“
NCV
”
button to enter the EF measurement.
Accuraacy
Range
6000counts
Resolution
6A
±(
2%+30
)
0.001A
10A
0.01A