AlphaTackPlus instrument case
The instrument is made of stainless steel or anodized aluminum parts only that are usually
compatible with standard cleaning solvents like acetone or IPA. Take care not to clean the
elastomer roller with it. The force sensor within the metal roller is made of ceramics and is
compatible to standard cleaning solvents also. The display foil on front side can be cleaned
with the cleaning agent
Böttcherin blau
.
6.3.
Cleaning agents
For best tack reproducibility the cleaning agents are not allowed to have non-evaporating
ingredients like surfactants or salts.
For conventional roller cleaning it is highly recommended to use simply petroleum benzine.
For UV Inks a mixture of ethanol and ethyl acetate has been proved to be successful, like:
Wash Up W907, Art.-No.: W907-K042, 25 kgs can
Supplier:
Gmelin GmbH & Co. KG, Schlossstraße 20,73054 Eislingen, Germany
Phone: ++49 7161 802 0,
www.zeller-gmelin.de
WIPEX Special cleaning rag, white
Viscose-Fleece-Rag, approx. 75g/mý, 29 x 38 cms
Solvent-consistent, 400 rags
Product-No.: 1504305
Supplier:
FINCKE-Hygiene Fachgroßhandel OHG,In der Florinskaul 8, 56218 Mülheim-Kärlich,
Germany
Phone: ++49 261-2 40 41,
www.fincke-hygiene.de