NovoPen 3 PenMate Instruction Manual
Page 36
FDA revision #4 (final)
§
The NovoPen 3 PenMate is not recommended for the blind or visually impaired,
487
without the assistance of a sighted individual who knows how to use NovoPen 3
488
PenMate.
489
§
If you use more than one type (N, R, or 70/30) of insulin, use a separate NovoPen 3
490
PenMate for each type of insulin.
491
§
The American Diabetes Association recommends that insulin should be self-
492
administered. The proper age for initiating this should be assessed by the adult
493
caregiver.
494
§
Use only a new PenFill cartridge when loading the NovoPen 3 PenMate. Never load
495
the NovoPen 3 PenMate with a partially filled PenFill cartridge.
496
§
The NovoPen 3 PenMate is designed for use with PenFill 3 mL insulin cartridges and
497
NovoFine single-use disposable needles.
498
§
Novo Nordisk is not responsible for any consequences arising from the use of the
499
NovoPen 3 PenMate with products that are not recommended by Novo Nordisk.
500
IMPORTANT (cont.)
501
Guidelines for storing the NovoPen 3 PenMate and PenFill 3 mL cartridges:
502
§
Store the NovoPen 3 PenMate (with the PenFill cartridge inside) at room
503
temperature.
Do not store
the NovoPen 3 PenMate in a refrigerator or areas where
504
there may be extreme temperatures or moisture, such as your car. Once the cartridge
505
is punctured, it can be kept at room temperature for the length of time identified in
506
the storage information section of the Information For The Patient supplied with the
507
PenFill cartridges.
508
§
Store the NovoPen 3 PenMate
without
the NovoFine needle attached and
with
the
509
pen cap in position. Leaving the needle on can cause the insulin to leak. This will
510
affect the concentration of an insulin suspension.
511
§
For information on storing PenFill cartridges, see the package insert that comes in
512
the PenFill cartridge box. Once the cartridge is punctured, it can be kept at room
513
temperature for the length of time identified in the storage information section of the
514
Information For The Patient supplied with the PenFill cartridges. The
expiration
515
date on the cartridge is for unused cartridges under refrigeration.
516
Содержание NovoPen 3 PenMate
Страница 1: ...1 ...
Страница 4: ...NovoPen 3 PenMate Instruction Manual Page 4 FDA revision 4 final 51 ...
Страница 5: ...NovoPen 3 PenMate Instruction Manual Page 5 FDA revision 4 final 52 PenFill Holder ...
Страница 18: ...NovoPen 3 PenMate Instruction Manual Page 18 FDA revision 4 final 272 273 274 275 ...
Страница 20: ...NovoPen 3 PenMate Instruction Manual Page 20 FDA revision 4 final 289 ...
Страница 22: ...NovoPen 3 PenMate Instruction Manual Page 22 FDA revision 4 final 302 ...
Страница 29: ...NovoPen 3 PenMate Instruction Manual Page 29 FDA revision 4 final 390 391 392 ...
Страница 34: ...NovoPen 3 PenMate Instruction Manual Page 34 FDA revision 4 final 448 ...