User Guide ProCair Prime

Protection Against Hazards

Fluids

Avoid fluid contact on any part of the control handset 
or control unit, by keeping mattress top cover zipped 
closed and the waterfall skirt down. If spills do occur:

 

Turn off control unit power and disconnect the 
unit from mains electricity supply.

 

Immediately clean fluids from the casing by 
wiping with a soft cloth.

  Ensure there is no moisture in or near the power 

inlet, control handset and power cord before 
reconnecting the power supply.

  Check the operation of controls and other 

components around the spill area.

  Fluid or liquid remaining on the electronic 

controls can cause corrosion that may cause 
the electronic components to fail. Component 
failures may cause the unit to operate erratically, 
possibly producing potential hazards to patient 
and carers.

Explosion Hazard

Equipment is not suitable for use in the presence of 
a flammable anaesthetic mixture with air, oxygen or 
nitrous oxide.

 

Do not use in the presence of smoking materials 
or open flame – air flowing through the mattress 
will support combustion.

 

Do not open the control unit – risk of electrical 
shock. Refer servicing to qualified service 
personnel.

 

Disposal

Dispose of all components (control unit including 
batteries, air filter, air cells, mattress cover and base) 
according to local procedures and regulations or 
contact Novis Healthcare for advice.

Power Cord

Periodically inspect the power cord for damage. 
The system should never be operated with a worn 
or damaged power cord. Keep the cord away from 
heated surfaces. Should the power cord be found to 
be worn or damaged, contact Novis Healthcare for a 
replacement.

Interference

Although this equipment conforms to the 
intent of directive IEC 60601-1-2

1

 in relation to 

Electromagnetic Compatibility, all electrical 
equipment may produce interference. If interference 
is suspected, move equipment away from sensitive 
devices or contact Novis Healthcare. 

1

 

 IEC 60601-1-2. Medical Electrical Equipment - Part 1: 
General Equipments for Safety, Amendment No. 2. 
Collateral Standard. Electromagnetic Compatibility 
Requirements and Test).

Safety Precautions

Содержание ProCair Prime

Страница 1: ...U S E R G U I D E ProCairPrime...

Страница 2: ...product improvement and reserves the right to amend specifications presented in this guide Information correct at time of production May2018 2018 Novis Healthcare All rights reserved Important Notice...

Страница 3: ...tions 7 8 System Preparation 9 12 Patient Set Up 13 Operation Control Handset 14 Operation Modes of Operation 15 Operation Quick Twist CPR 15 Operation Transport Mode 16 Operation Deflation Storage 17...

Страница 4: ...lood and lymphatic fluids through these tissues to provide essential oxygen and remove metabolic waste A full size movable foam sheet can be positioned at the bottom of the mattress working in conjunc...

Страница 5: ...Carry bag It is recommended that all packing materials and the User Guide be kept in the carry bag provided for ease of storage and or transport HEAD END FOOT END FOOT END HEAD END RIGHT RIGHT LEFT LE...

Страница 6: ...fessionals of sound mind with the ability to read and understand English and Westernised Arabic Numerals Working Environment Temperature 15 C to 35 C 59 F to 95 F Humidity 30 to 75 non condensing Ship...

Страница 7: ...cell movement As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams objects in pockets etc Never use sharp obje...

Страница 8: ...o not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the control unit risk of electrical shock Refer servicing to qualified...

Страница 9: ...nd On a profiling bed secure the straps around the moveable sections of the base Ensure the buckles are securely fastened and straps are pulled tight Do not secure mattress straps to bed side rails st...

Страница 10: ...connectors are secure by pushing the male and female ends together there should be no movement If a connection is open a click will be heard once connector is firmly closed System Preparation 6Feed p...

Страница 11: ...ts to pressure setting 02 upon full initial inflation Press AUTO SET to activate automatic pressure setting This feature monitors patient weight every alternation cycle and automatically selects the a...

Страница 12: ...onvert to hybrid mode 1 Unzip zippers from the foot end of the mattress and remove the top cover 2Undo 4 press studs on one side of the mattress base cover and remove the foam sheet from its compartme...

Страница 13: ...deflated You may need to unzip the cover to feel the cells for inflation 2With the patient lying supine unzip one side of the top cover just past sacral region lower spine 3Slide your hand underneath...

Страница 14: ...tton the system will automatically detect patient weight every alternating cycle and adjust pressure accordingly if necessary E Manual Pressure Select After initial inflation the system defaults to se...

Страница 15: ...tage the ProCair Prime Universal Therapy System will stop functioning and PF arlam code will be displayed on the Alarm Indicator accompanied by an audible alarm The system will return to its normal op...

Страница 16: ...t to the CLOSED position Switch off control unit by long pressing the Power button for 3 secs Open top cover disconnect the air connectors from the control unit and connect the male connector to the f...

Страница 17: ...from the system detach the mattress from the bed by unfastening all mattress straps place handset and power cord on the centre of the mattress then fold and roll the mattress from head end to foot end...

Страница 18: ...d Foam Sheet Cleaning Swab the mattress base air cells and the foam sheet with a solution of sodium hypochlorite or similar up to 10 000 ppm available chlorine Dry thoroughly before reassembly Do not...

Страница 19: ...her zipper run the zipper along the side of the mattress towards the centre of the head end 3 Repeat with other zipper The top cover can now be detached from the mattress base To reattach the top cove...

Страница 20: ...ply Check for loose connection on plug and main power is switched on Check condition of power cord and plug Check if mains socket is faulty Control unit does not operate no display or lights Check con...

Страница 21: ...2 0 1A MIN MAX PRESSURE 20 60 mmHg 6 mmHg PROTECTION TYPE Class II Type BF INGRESS PROTECTION RATING IP22 HANDSET MODEL HS NBE HANDSET PART NO FO NVS0001 HANDSET CASING MATERIAL ABS MATTRESS DIMENSION...

Страница 22: ...onsible and contact your local authority on available options to recycle this product at its end of life Service Life The expected service life of a control unit and a mattress is highly dependent on...

Страница 23: ...ation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the...

Страница 24: ...60 Hz 50 Hz NOTE UT is the AC mains voltage prior to application of the test level Conducted RF IEC 61000 4 6 3 Vrms 0 15 MHz 80 MHz 3 Vrms 0 15 MHz 80 MHz Portable and mobile RF communications equipm...

Страница 25: ...Pulse modulation B 217 Hz 0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation B 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900...

Страница 26: ...aired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for any repair other t...

Страница 27: ...27 ProCair Prime User Guide Notes...

Страница 28: ...Pressure care and patient handling specialists novis com au 1300 738 885...

Отзывы: