14
User Guide
CairMax Duo
Care and Cleaning
To prevent cross contamination, the mattress should be examined and disinfected between patient use.
Clean the mattress in accordance with local infection control policy and government regulations.
Failure to do so could cause patient or personal injury.
The mattress is not protected against excessive amounts of fluid.
Switch off and disconnect the control unit from mains power supply before cleaning.
Disconnect power supply before cleaning. Do not immerse the control unit in fluid.
Do not use high temperature autoclave steam cleaning devices or phenolic based products for cleaning.
This could result in damage to the equipment and may result in damage to the polyurethane coating, or
negate the biocompatibility properties of the fabric.
Cleaning and Infection Control
It is recommended that the CairMax Duo system is cleaned every two weeks if in constant use.
Top Cover Cleaning
Unzip and remove the top cover from the base before
washing (refer below for instructions). For basic care
and cleaning, wipe down with warm water containing
neutral detergent. The top cover can also be machine
washed at 95° C using neutral detergents.
Refer to the top cover wash tag for detailed
cleaning instructions.
Do not use system without top cover.
Top Cover Removal
1
Raise the waterfall skirt and locate the zippers at
the foot end of the mattress.
2
Run each zipper along the side of the mattress
from foot end to head end.
3
Separate top cover from mattress base.