Novis CairMax Скачать руководство пользователя страница 1

U S E R   G U I D E

CairMax Duo

Содержание CairMax

Страница 1: ...U S E R G U I D E CairMaxDuo...

Страница 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Страница 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Страница 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Страница 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Страница 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Страница 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Страница 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Страница 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Страница 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Страница 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Страница 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Страница 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Страница 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Страница 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Страница 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Страница 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Страница 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Страница 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Страница 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...