ProCair Trio User Guide

Troubleshooting

Alarm Codes

A set of 3 amber lights will be lit in 5 diferent patterns, accompanied by an audible alarm and accompanying 
Alarm Code display, to indicate the control unit or mattress is experiencing a fault. There are also additional 
warning lights, accompanied by audible beeping alarm, to alert the user immediate recti

fi

  cation is needed.

The lights will continue to 

fl

  ash until the fault is cleared. The audible alarm can be silenced for 20 minutes 

by pressing the Alarm Mute button. It will reactivate if the fault is not recti

fi

  ed , or if a new fault is detected.

CODE AND TRIGGER

SOLUTION

Air cells have failed to reach the pre-set pressure

Ensure the main power is turned on and power cord is connected to 
mains and control unit.

Check control unit/mattress air connections are 

fi

 tted securely, 

and reconnect umbilical cord if loose. 

Ensure CPR valve is set to CLOSE position. Replace CPR valve if air 
leak is found. 

Check air intake from 

fi

  lter is not blocked by linen/dust. Replace 

with new 

fi

  lter if needed.

Air cell pressure exceeds the pre-set pressure

Check the air hoses for kinks, obstructions or damage. Undo any

kinks and obstructions.

Air cells have failed to reach operating pressure

after turning on

Turn of control unit and disconnect from power. Reconnect to 
power after 1 minute and restart initiation process.

Check any leaks on control unit/mattress air connections, CPR 
valve and air cells.

Air cells have failed to alternate

Remove patient from mattress. Turn of control unit and disconnect 
from power. Reconnect to power after 1 minute and restart 
initiation process, return patient to mattress once initiated and 
alternation has resumed.

If issue persists, contact Novis Healthcare. A service may be 
required.

Control unit has failed to detect connection to patient 

weight sensor and exit pad sensor

Ensure air hose connector is securely fastened and reconnect if 
loose.

Bed Exit pad has detected patient egress.

Check for patient on mattress.

To reset bed exit alert, turn off bed exit alarm by pressing the 
“Bed Exit” button, return patient to mattress, press “Bed Exit” again 
to reactivate alert.

Battery charge below 10% of capacity

Check the control unit is connected to the mains power supply and 
the power is operational.

LOW PRESSURE

HIGH PRESSURE

STARTUP FAILURE

ALTERNATION FAILURE

SENSOR DISCONNECTED

BED EXIT ALARM

FLASHING: LOW BATT

Содержание APMPC-R05

Страница 1: ...pressure care and patient handling specialists novis com au 1300 738 885 Static Mode Programming Option ProCairTrio UserGuide...

Страница 2: ...dreservestherightto amendspecificationspresentedinthisguide Informationcorrectattimeofproduction September2020 2020NovisHealthcare Allrightsreserved Definition of Symbols Used Thefollowingsymbolsmayap...

Страница 3: ...tion 9 Patient Set Up 12 Operation Control Panel 13 Operation Mode 16 Operation Pull Cap CPR Release 17 Operation Deflation Storage 18 Care and Cleaning 19 Troubleshooting 21 Technical Specifications...

Страница 4: ...atient ssurface tissue maintainingandstimulatingtheflowof bloodandlymphaticfluidsthroughthesetissues toprovideessentialoxygenandremovemetabolic waste Thesystemconsistsofthefollowingcomponents Mattress...

Страница 5: ...areHomehealthcareand Professionalhealthcare Working Environment ControlUnit Temperature 15 Cto35 C 59 Fto95 F Humidity 30 to75 non condensing Shipping Storage Environment ControlUnit Temperature 5 Cto...

Страница 6: ...ndividuals Thisdevicecanbeusedinhomehealthcareand professionalhealthcareenvironment Medicalelectricalequipmentneedsspecial precautionsregardingEMCandneedstobe installedaccordingtotheEMCinformation pro...

Страница 7: ...ely Avoidblockingtheairintakesofthecontrol unit locatedattherearoftheunit Donotplace itemssuchasblanketsoverthecontrolunit Bedframesusedwiththesystemscanvary greatlydependingonthespecifichealthcare se...

Страница 8: ...fsmoking materialsoropenflame airflowingthrough themattresswillsupportcombustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified servicepersonnel Disposal Disposeofallcomponen...

Страница 9: ...wardsthebaseofthebed 2 Attachto the bed bysecuringthe adjustable straps locatedontheundersideofthe mattressbaseundereachbedend Ona profilingbed securethestrapsaroundthe moveablesectionsofthebase Ensur...

Страница 10: ...dhanging hooks Pullthehooksbytherubbertabsto preventaccidentallytrappingyourfingers Ensureitissecurebeforeuse failuretodoso couldresultinequipmentdamage 6 Connecttheumbilicalconnectorstothe socketsons...

Страница 11: ...flash green Allow upto 20minutes for completeinflation Donotlieaperson orany weight onthemattressduringinitialinflation Donotplaceaperson oranyitem onthe mattressduringinitialinflation 9 Wheninitialin...

Страница 12: ...to feelthe cellsforinflation 2 Withthepatientlyingsupine unzipone sideofthe top cover justpastsacralregion lowerspine 3 Slideyourhandunderneaththepatientand feelfor a deflatedcellunderthepatient s lo...

Страница 13: ...therapyanduseradjustablepressuresettings TheselectedmodeisindicatedbyanLEDindicator D AlarmIndicator ThethreeamberLEDslightupin5different patternstoindicatedifferentfaultsthesystem maybeexperiencing...

Страница 14: ...Thisfeaturehelpstopreventtheriskofbottoming out Agreen AUTOFOWLER lightilluminateswhenthis functionisactive Itisautomaticallydeactivated whenthemattressbackrestisprofiledbelow5 HPower StandbyandServic...

Страница 15: ...ruptionto therapy due to power failure Note Itwilltake16 20hourstofullychargethe battery K BedExitAlarm PressingthebuttonturnsBedExitAlarmonandoff TheONlightabovebuttonilluminatestoindicatethe functio...

Страница 16: ...icallydisengage Backup Battery Thelithium ionbackupbatteryisdesignedto providetemporaryuninterruptedtherapywhen mainspowerisunavailableegduringpatient transferorinapoweroutage Thecontrolunitautomatica...

Страница 17: ...ksareheard Airwillremaininthe systemforupto24hours dependingonpatient andenvironmentalcircumstances Specialised Heel Therapy ThespecialisedHeelCairzoneenablestheseven narrowalternatingaircellsinthehee...

Страница 18: ...sedfromthesystem detachthemattressfromthebedby unfasteningthestraps thenfoldandroll themattressfromheadendtofootendfor storage 6 Returnallitemstothecustomcarrybagfor safekeeping CPRRELEASE POWER STAND...

Страница 19: ...hadryclothtoremoveexcess moisture Asoftbristlednylonbrushcanbeusedto gentlycleancrevices Ensurethecontrolunitisdisconnectedfrom mainspowerbeforecleaning Donotspraydisinfectantdirectlyontothe controlun...

Страница 20: ...lunitfacedownonasoft flat surfacewithbackpaneluppermost useasoft clothtopreventscratches 3 Useasmallscrewdrivertocarefullyremovethe airfiltercover Cleanthedustfromthefilteror discardandreplacewithanew...

Страница 21: ...for kinks obstructions or damage Undo any kinks and obstructions Air cells have failed to reach operating pressure after turning on Turn of control unit and disconnect from power Reconnect to power a...

Страница 22: ...ate no display lights Check control unit is connected to the mains power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if ne...

Страница 23: ...ENT 0 4 0 2A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg FLOW RATE Max10L min PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 CONTROL UNIT DIMENSION HEIGHT 326mm WIDTH 275mm DEP...

Страница 24: ...dsThefirst digit indicates thelevelofprotectionthattheenclosure providesagainst accessto hazardous parts e g electricalconductors moving parts andtheingressofsolidforeignobjects SecondDigit LiquidsPro...

Страница 25: ...nsequipment Fortransmittersratedatamaximumoutputpowernotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherePisthe maxi...

Страница 26: ...ative humidity should be at least 30 Air 2 4 8 15 kV Air 2 4 8 15 kV Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality should be t...

Страница 27: ...eryInUse LEDisnormal duetopowerissuppliedbyinternalbatteryduringpoweroutage Thebackupbatterycouldbeusedforupto3hours EMI EMC Statement and Manufacturer s Declaration Immunitytest IEC60601testlevel Com...

Страница 28: ...mendedseparationdistance calculatedfromtheequationapplicable tothefrequencyofthetransmitter Radiated RF IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AMat1kHz 10V m 80MHz 2 7GHz 80 AMat1kHz Recommendedseparatio...

Страница 29: ...on Pleasedonotuseanyothercablesoraccessoriesnotapprovedbythemanufacturerinthismanualtoavoidnegative influenceonelectromagneticcompatibility EMI EMC Statement and Manufacturer s Declaration Testfrequen...

Страница 30: ...ished good of the same type rather than being repaired Refurbished parts may be used to repair goods NovisHealthcarewillnotbeheldresponsibleforanyrepair otherthanthosecarriedoutbyitoroneofitsAuthorise...

Страница 31: ...31 ProCair Trio User Guide Notes...

Страница 32: ...Pressure care and patient handling specialists 06102020F 1300738885 novis novis com au 12 12MarsRd LaneCoveWest NSW2066...

Отзывы: