21
ProCair Trio User Guide
Troubleshooting
Alarm Codes
A set of 3 amber lights will be lit in 5 diferent patterns, accompanied by an audible alarm and accompanying
Alarm Code display, to indicate the control unit or mattress is experiencing a fault. There are also additional
warning lights, accompanied by audible beeping alarm, to alert the user immediate recti
fi
cation is needed.
The lights will continue to
fl
ash until the fault is cleared. The audible alarm can be silenced for 20 minutes
by pressing the Alarm Mute button. It will reactivate if the fault is not recti
fi
ed , or if a new fault is detected.
CODE AND TRIGGER
SOLUTION
Air cells have failed to reach the pre-set pressure
Ensure the main power is turned on and power cord is connected to
mains and control unit.
Check control unit/mattress air connections are
fi
tted securely,
and reconnect umbilical cord if loose.
Ensure CPR valve is set to CLOSE position. Replace CPR valve if air
leak is found.
Check air intake from
fi
lter is not blocked by linen/dust. Replace
with new
fi
lter if needed.
Air cell pressure exceeds the pre-set pressure
Check the air hoses for kinks, obstructions or damage. Undo any
kinks and obstructions.
Air cells have failed to reach operating pressure
after turning on
Turn of control unit and disconnect from power. Reconnect to
power after 1 minute and restart initiation process.
Check any leaks on control unit/mattress air connections, CPR
valve and air cells.
Air cells have failed to alternate
Remove patient from mattress. Turn of control unit and disconnect
from power. Reconnect to power after 1 minute and restart
initiation process, return patient to mattress once initiated and
alternation has resumed.
If issue persists, contact Novis Healthcare. A service may be
required.
Control unit has failed to detect connection to patient
weight sensor and exit pad sensor
Ensure air hose connector is securely fastened and reconnect if
loose.
Bed Exit pad has detected patient egress.
Check for patient on mattress.
To reset bed exit alert, turn off bed exit alarm by pressing the
“Bed Exit” button, return patient to mattress, press “Bed Exit” again
to reactivate alert.
Battery charge below 10% of capacity
Check the control unit is connected to the mains power supply and
the power is operational.
LOW PRESSURE
HIGH PRESSURE
STARTUP FAILURE
ALTERNATION FAILURE
SENSOR DISCONNECTED
BED EXIT ALARM
FLASHING: LOW BATT