TIP
The ARGES system controller has several interfaces of connector type RJ45.
When connecting the cables pay attention to the labeling on the housing.
4.
Preparing for use
4.3. Installing
3. Check for leakage by carefully turning on the purge gas and watching the connection.
If purge gas is leaking then immediately stop the procedure, disconnect the purge
gas, see the procedure further below, and redo the connection.
Remark: The purge gas outlet is around the protective window of the laser aperture.
4. Turn off the purge gas until the superordinate machine is switched on.
How to disconnect the purge gas
1. Turn off the purge gas supply line.
2. Relieve the pressure from the hose.
3. Press the end piece (marked red in figure 4.11 on page 25) into the quick connector,
hold it pressed down and pull out the hose.
Figure 4.11.: End piece of the purge gas quick connector (marked red)
4.3.5. Connecting the ARGES system controller
Control the scan head via an ARGES system controller.
Target audience and qualification
Experts
Procedure