CONNECTING
PAGE
|
22
Connecting
The following table provides dew point temperatures for a range of air temperature and relative humidity
values. Remember that the laser’s coolant temperature must be set
above
the dew point temperatures
given in the chart; however, for best results and performance, use a coolant temperature in the range of
18–22 °C (64–72 °F) when possible.
Table 2- Dew point temperatures.