NovaLynx Corporation 
 

 

255-704-B 

Page 12 

April 2021 

 

 

Raw 

Scaled 

Units: 

Volts 

inch / mm 

Value 1: 

 

  

Value 2:    

  

 
 

1.

 

Value 1: 

a.

 

Fill the evaporation pan with approximately 5 inches of water. Allow the water to settle. 

b.

 

Observe the Number 1 Voltage measurement shown in the Current Readings window. 
Record the voltage. 

c.

 

As accurately as possible, measure the depth of water in the pan. Record the level. 

2.

 

Value 2: 

a.

 

Add water to the evaporation pan until the water is approximately 8 inches deep. Allow 
the water to settle. 

b.

 

Record the Number 1 Voltage measurement. 

c.

 

As accurately as possible, measure the depth of water in the pan. Record the level. 

3.

 

Close the Status window. 

4.

 

Open the Launch Logger window and navigate to the Linear Scaling Assistant window 
(Section 5.4). Enter your measurements in place of the generic parameters. 
 

 

5.

 

Click "Update" to save your measurements. 

6.

 

Return to the Status screen and compare the calibrated result (Number 5 Ev Pan Level) with the 
Value 2 depth measurement. They should be very close, provided the water level has not 
changed. 

 

6

 

OPERATION 

6.1

 

Start Logging 

Connect your computer to the logger (Section 5.3). Click the "Launch device" icon to open the logger 
setup window. When ready, click the "Start" button (lower right) to start logging. The LCD display will 
now show the current position of the potentiometer expressed in inches (or millimeters). Look at the 
lower left corner of the LCD display and verify that the logger is "LOGGING". Disconnect the USB cable. 
 

Содержание 255-704-B

Страница 1: ...04 B UM 20210419 255 704 B User Manual 255 704 B Evaporation Logger 255 704 B Evaporation Logger with 255 100 Analog Output Evaporation Gauge Phone 530 823 7185 Email nova novalynx com Website www nov...

Страница 2: ...items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside of...

Страница 3: ...tions 5 5 2 Float and Chain Installation 6 5 3 Software Installation 6 5 4 Logger Settings 8 5 5 Offset Calibration optional 11 5 6 Field Calibration optional 11 6 OPERATION 12 6 1 Start Logging 12 6...

Страница 4: ...the current water level in the evaporation pan along with logging status battery status and memory consumption The logger stores up to 1 9 million measurements with 16 bit resolution The logger is pro...

Страница 5: ...is straightforward The enclosure is fitted with a large band clamp which must be opened and passed around the body of the evaporation gauge photo Adjust the height so that the logger does not interfer...

Страница 6: ...o clock position 30 degrees to left 3 Lift the chain until all the slack is taken up but the float is still resting on the bottom of the gauge Drape the chain over the sprocket so that the float is o...

Страница 7: ...to the computer 4 Look at the status ribbon at the bottom of the HOBOware program window below At first the status will indicate No devices connected Within a minute the status should update and show...

Страница 8: ...Logger has been pre programmed to read Channel 1 as water level in the evaporation pan The logger will be programmed either in inches or millimeters depending on the order specification The example be...

Страница 9: ...ant window will open Please verify the settings on your logger NOTE The generic calibration shown above does not take into account the amount of water required to raise the float off the bottom of the...

Страница 10: ...dow 11 Click the Device status icon to open the Status window below Note the battery level and memory used to ensure there is enough power and memory available for your next logging session 12 Set the...

Страница 11: ...er s reading at that level 1 Measure the water depth and calculate the offset Example 8 00 Measured Depth 6 30 Logger Reading 1 7 Offset 2 Add the offset to the generic calibration values Value 1 Scal...

Страница 12: ...ose the Status window 4 Open the Launch Logger window and navigate to the Linear Scaling Assistant window Section 5 4 Enter your measurements in place of the generic parameters 5 Click Update to save...

Страница 13: ...file location and filename then click Save Once the file is saved the Plot Setup screen will open Select the elements you want to display in graphical and tabular form then click Plot LEFT Plot Setup...

Страница 14: ...ust be deducted from the evaporation totals Water added by precipitation should also be deducted 7 MAINTENANCE At regular intervals inspect the following Check under the evaporation pan for leaks Insp...

Страница 15: ...it back into place The two AAA 1 5V batteries should provide you with 7 to 9 months of continuous use Note A desiccant pack may be inside the logger box It can be replaced or retreated as needed 9 TEM...

Страница 16: ...eight Drape the counterweight and chain over the outside of the gauge Do not attempt to remove the float 3 Rotate the sprocket by hand to verify that it turns freely 360 without binding It should spin...

Страница 17: ...the sprocket may have slipped on the shaft and be out of position If the readings on the display are progressively worse the farther the sprocket is turned the potentiometer may be out of specificatio...

Отзывы: