APPENDIX A 

 

255-100C-4 WIRING DIAGRAM

JDC

09-20-2018

255-100C-4

Calibration Procedure

890-0020-01

1 2

White - Signal

Black - Reference

Shield - Earth ground

Red - Excitation

Analog Output 

Evaporation Gauge

Model 255-100

mA   COM

R28        GAIN

 

R26        OFFSET

20°

19.1 mA

4.0 mA

Dead Band

1) Connect a 9V or 12V battery and meter as shown.

2) Rotate the gear until the lowest mA reading is obtained. Adjust R26 OFFSET to 4.00 ±0.080 mA.

3) Rotate the gear counter-clockwise. The output will rise gradually until the dead band

     is reached and then jump suddenly to 26-28 mA. Back off slightly to get out of the dead band 

     and adjust R28 GAIN to 19.1 ±0.080 mA.

4) There is a small interaction between the GAIN and OFFSET pots, so verify the lowest mA reading

     is still 4.00 ±0.080 mA.

 

Содержание 255-100C-4

Страница 1: ...nt Loop Signal Converter For use with the 255 100 Analog Output Evaporation Gauge And the 255 200 Evaporation Pan each sold separately 255 100C 4N Phone 530 823 7185 USA Toll Free 800 321 3577 Fax 530...

Страница 2: ...If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside of the bo...

Страница 3: ...on 200 100C 4 Page 3 September 2018 CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 TECHNICAL SPECIFICATION 4 4 INSTALLATION 5 5 WIRING DIAGRAM 7 6 CALIBRATION 7 7 SINGLE POINT OFFSET ADJUSTMENT 8 8 MAINTENAN...

Страница 4: ...hased as a bare board Part 255 100C 4 or enclosed in a weather tight IP65 enclosure Part 255 100C 4N Connections to the circuit board are via screw terminals The enclosure has two cable grip fittings...

Страница 5: ...r mounting to a mast up to 1 3 2 cm in diameter The mast mounting hardware can be removed and the enclosure mounted to a flat surface instead if required There are four mounting holes accessible when...

Страница 6: ...NovaLynx Corporation 200 100C 4 Page 6 September 2018 0 185 3 800 3 500 0 150 1 125 2 250 Figure 1 Figure 2...

Страница 7: ...he output will be meaningless Cable for connecting the 4 20 mA loop is not supplied Use a wire gage adequate to provide a minimum of 8 volts at the Signal Conditioner after voltage drop over the cable...

Страница 8: ...tooth raises or lowers the float 0 25 6 35 mm If finer adjustment is needed it may be possible to program the offset in your monitoring equipment 8 MAINTENANCE The 200 100C 4 Signal Conditioner requir...

Страница 9: ...nect a 9V or 12V battery and meter as shown 2 Rotate the gear until the lowest mA reading is obtained Adjust R26 OFFSET to 4 00 0 080 mA 3 Rotate the gear counter clockwise The output will rise gradua...

Отзывы: