Nova APP-2001 Скачать руководство пользователя страница 10

10

MAINTENANCE

 

Ø

General

•  Check main power cord and plug if there are abrasions or excessive 

wear 

•  Check mattress cover for signs of wear or damage  Ensure mattress 

cover and tubes are connected together correctly 

•  Plug in the pump unit and check the airflow from the hose 

connection port  The airflow should alternate between ports every 
half-cycle time 

•  Check the air hoses if there are kinks or breaks  For replacement, 

please contact the dealer from whom you purchased 

•  Make sure the mattress tube is well connected 
•  Check the pump unit and make sure both power and power 

indicator are off when the switch is turned off 

  NOTE: In case the fuse of the pump needs to be changed, use a 

1-Ampere fuse.

 

Ø

Low pressure

To examine if there is air leakage between pump and mattress pad 
connections or from air mattress tubes:

•  Check connectors between air mattress pad and pump  If they are 

disconnected, please reconnect them 

•  Check air-connecting tubes  Ensure each single cell is not broken  

Set pressure dial to firm position  Keep the tubes fully inflated and 
inspect for air leakage 

•  Exam air leakage from cells  Ensure no leakage occurs  If any 

leakage occurs, please contact  

 

Ø

Changing air filters

It is recommended to change the air filter on the pump every 6 
months of continued use  Please contact the dealer from whom you 
purchased for replacement filters 

Содержание APP-2001

Страница 1: ...DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE NOVA Pump Pad On Off Switch Pressure Adjustable Item APP 2001...

Страница 2: ...is necessary when this product is used by on or near children or invalids Use this product only for its intended use as described in this manual Do not use attachments not recommended by the manufactu...

Страница 3: ...ssure set up 7 Low pressure 8 CLEANING Pump 8 Mattress 8 HANDLING AND STORAGE Storage instructions 9 MAINTENANCE General 10 Low pressure 10 Changing air filters 10 TECHNICAL DESCRIPTION America Canada...

Страница 4: ...and certificated to the following standards UL 1431 120V 60Hz FDA 120V 60Hz Intended use NOVA series is intended to reduce the incidence of pressure ulcers while optimizing patient comfort It also pr...

Страница 5: ...filter kit a c b d 1 Position pump on horizontal surface or hang it on the frame of the foot end with the hooks Ensure the pump is stable DO NOT operate pump on carpet 2 Open mattress pad flat on the...

Страница 6: ...ource and switch on The switch indicator will light up When switched off the indicator will turn off 5 For the first inflation to save inflating time adjust the pressure adjust knob to FIRM setting Al...

Страница 7: ...d and allow pad to expel air completely Pressure set up Users can adjust the pressure of the air mattress pad to a desired firmness by themselves or by the suggestion of a health care professional NOT...

Страница 8: ...e and spray or wipe the outside of the case with quaternary ammonium solution DO NOT spray any cleaning solution directly on the surface of the pump DO NOT use a hypocarbonate or phenolic based cleani...

Страница 9: ...om one side and are sprayed on all sides with a disinfectant Let it sit for the required incubation time and wipe down with a clean cloth Make sure to disconnect all the air cells one by one and spray...

Страница 10: ...ndicator are off when the switch is turned off NOTE In case the fuse of the pump needs to be changed use a 1 Ampere fuse Low pressure To examine if there is air leakage between pump and mattress pad c...

Страница 11: ...thead screwdriver gently pry up on the edge of the filter cap 3 Remove dirty filter and insert a clean filter into the filter cap Place filter cap into pump TECHNICAL DESCRIPTION NOVA pump mattress pa...

Страница 12: ...tion Environment Temperature Humidity Operation 50 F 95 F Storage 59 F 122 F Shipping 59 F 158 F Operation 20 80 non condensing Storage 10 90 non condensing NOTE The specifications also apply to those...

Страница 13: ...of electric shock by providing an escape wire for the electric current This product is equipped with a cord having a grounding wire with a grounding plug The plug must be plugged into an outlet that...

Страница 14: ...alled by a qualified electrician If the product must be reconnected for use on a different type of electric circuit the reconnection should be made by qualified service personnel Extension Cords If it...

Страница 15: ...se modified without prior written authorization are not covered within this warranty For warranty service please contact the dealer from whom you purchased your NOVA product In the event that you do n...

Страница 16: ...NOVA Medical Products 1470 Beachey Place Carson CA 90746 USA MADE IN CHINA 4FL1 E346315...

Отзывы: