Interfering Substances
Lactate Interferences:
The Nova StatStrip Xpress Lactate Hospital Meter exhibits
no
interference from the following substances up to the
following concentration levels:
Substances
Conc
mg/dL(mmol/L)
Acetaminophen
20.0 (0.66)
Ascorbic Acid
10.0 (0.57)
Bilirubin
15.0 (0.26)
Cholesterol
500.0 (12.9)
Creatinine
6.0
(0.53)
Dopamine
10.0 (0.53)
Ephedrine
0.9
(0.055)
Glucose
900.0 (49.7)
Ibuprofen
48.0 (2.33)
Substances
Conc
mg/dL(mmol/L)
L-Dopa
5.0
(0,25)
Methyl-Dopa
1.0
(0.042)
Salicylate
30.0 (1.87)
Tetracycline
30.0 (0.62)
Tolazamide
15.0 (0.48)
Tolbutamide
45.0 (1.67)
Triglycerides
750.0 (8.78)
Uric Acid
20.0 (1.05)