E3 Used Strip Error
The test strip was previously used.
Action:
Repeat the test with a new test strip.
If the error persists, perform the test using an
alternate test strip vial or alternate method.
E4 Short Sample Error
An insufficient sample volume (Control or blood)
was drawn into the test strip.
Action:
Repeat the test with a new test strip.
If the error persists, perform the test using an
alternate test strip vial or alternate method.