Cleaning, disinfection and sterilization
The following points in particular are important with regard to caring for the material:
●
Perform cleaning, disinfection and sterilization after every treatment!
●
Always autoclave the material in sterilization packaging.
●
Make sure that sterilization packaging is not more than 80 % filled.
●
Autoclave the material at 135°C for at least 5 minutes.
●
If sterilized material is not used immediately, the material packaging must be labeled with the
sterilization date.
●
Nouvag AG recommends including a sterility indicator.
8.1
Control unit and On/Off pedal
Control unit and On/Off pedal do not come into contact with the patient.
Wipe the outside using micro-biologically tested surface disinfectant or a 70 % isopropyl solution. Don’t use
spray cleaning and spray disinfection, the front plate of the control unit is not sealed for this purpose.
8.2
Micro motor Endo
The micro motor does not come into contact with the patient.
Wipe the outside using micro-biologically tested surface disinfectant or a 70 % isopropyl solution.