Please check that all  
expected items are  
included in the shipment: 

 

 The instrument  

 External power/signal cable 

 NORTEK equipment storage 

    box/toolkit 

 RS485 to USB high speed adaptor 

 AC/DC voltage transformer  

 Power cable  

    (European or American)  

 250 ml bottle with seeding  

 Software on USB memory stick 

 Final test check list 

 Packing list 

 Warranty card 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Install the Nortek software  

The instrument’s software is 
located on the attached 
memory stick.  
 
Please check the web 
regularly for updates on both 
software and firmware. 

 

 

This Quick Guide provides you with the basic information you need to get the 
Vectrino Profiler up and running as quickly as possible. The sensor is an 
extremely powerful tool for flow measurement. It is a seemingly simple system but 
requires consideration of many different parameters to produce the highest quality 
data. The biggest key to its successful use if an understanding of the basic 
concepts of its operation and the potential problems that can occur. Experience 
with the system is an important factor as well, and the value of test deployments 
cannot be overemphasized. We encourage you to consult the Vectrino Profiler 
User guide (available within the software) to get a more profound knowledge 
covering both the sensor- and software capabilities. You will also find lots of useful 
information at the Nortek web: 

www.nortekgroup.com

 

Preparation 

 

Install the software located on the USB memory stick. The Vectrino Profiler 
acquisition software requires Java virtual machine (JVM) to operate (download 
from www.java.com if not prompted by the software).  

 

Using the enclosed cables, plug in the AC adapter, and connect the Vectrino 
Profiler cable and USB converter to the USB port on your computer. At high 
sampling rates (above 60 Hz), the driver parameters need to be tuned to 
provide an adequate display rate for the application software. Without this 
tuning, higher sampling rates will result in a sporadic display update. Please 
refer to the 

SW User Guide

, in the software (click Help) for more information. 

 

Start the Vectrino Profiler software.  

 

Click 

Communication > Connect

 

>

 

Vectrino 

 

  Select the appropriate serial port and port speed (baud rate) and press 

Apply

It is recommended to use a 

baud rate of 937500

 for optimized performance. 

You may take this opportunity to provide a specific “Instrument Name” for your 
device. This may help differentiate instruments if you have several in the same 
experiment.  

  Run a Probe check in accordance with the procedure described in the 

Vectrino 

Profiler User Guide

 to verify that the instrument is working as expected.  

  Test the measurement quality with the Probe Check function: Open 

Configuration; 

tick the 

Probe Check 

and 

Start Data Collection

. Make sure 

the instrument is submerged in water during this test.

 

 

Configuration  

  Click the 

Configuration

 button to activate the dialog for configuring the 

instrument velocity settings, bottom check settings and disk file parameters. A 
description of each submenu and each parameter can be found in the 

SW User 

Guide. 

Extra attention should be given to the Velocity range parameter. Read 

more about Weak Spots below.

 

  When the parameters are adjusted, press 

Apply 

and/or 

OK 

and accept the 

changes. 

 

  By using the 

Open/Save

 commands in the File menu the configuration 

parameters can be saved to a *.nts file at any given time and reloaded when it 
is time to deploy the instrument

 

Getting Started 

  Data can be displayed by clicking the 

Start Collecting Data 

button. 

  After data collection has started, the button switches to 

Stop Collecting Data

The button switches back to Start mode when the instrument has been verified 
to complete collecting data.

 

  The display window shows profiling and time series graphs for data currently 

being collected. A status display is shown on the upper right of the panel. This 
shows four lights used to display status information while collecting data (run 
status, collection status, connection status, and instrument status). Hovering 
the mouse over each status light will display more detailed information. 

 

 

Save to Disk

 can be enabled either before data collection is started or during 

data collection. The preferred mode is enabling before starting collection.

 