Nortek Scour Monitor Скачать руководство пользователя страница 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stop Recorder Deployment

 

  Select 

Stop  Recorder  Deployment

.  When  data  acquisition  has  stopped  a 

dialog will display the current Scour Monitor and PC clock time. 

 

Data Retrieval 

  Select Recorder Data Retrieval. 

  The recorder overview dialog will show a list of all files currently stored in the 

recorder, including the time of the first measurement and the size for each 
file. 

  Select a file in the list by clicking on the filename. Press 

Retrieve

 to specify 

the name and where to save the file.  

 

Data Conversion 

  To convert binary data files retrieved from the recorder to a readable ASCII 

format select 

Data Conversion 

from the Deployment menu or press the 

Data Conversion toolbar button. 

  Add files to the 

Files to convert

 list by using the 

Add file...

 file selection 

dialog. To convert the files, select items in the list and press the blue arrow. 
Specify the location for the converted files in the Save in folder field. An 
optional pre- or postfix that will be added to the recorder filename may be 
specified. Select from the 

View files 

drop-down list to open the converted 

(ASCII) files in Notepad. 

 

Tip! Check the retrieved data before deploying the instrument for the second 
time. If something is wrong with the instrument, the deployment or the 
configuration, this will be detected and the possibility of making the same 
mistake twice is avoided. 

 

On-line Data Collection 

  An on-line data collection module is included for test purposes and to provide 

data collection from the Scour Monitor in real time. The program will also 
store data to binary files on the computer hard disk for post processing and 
analysis. 

  Select 

Start Data Collection

 from the On-line menu or press the toolbar 

button to start data acquisition. The real time display shows velocity data and 
sensor data as it is collected. 

  To capture the data that comes in over the serial port to disk select 

Disk 

Recording…

 from the On-line menu and define the name of a file (without 

extension). Once this disk file is defined, the Start Disk Recording and Stop 
Disk Recording menu items and toolbar buttons will be enabled. Once data 
recording to disk has started, the status box on the display screen will show 
the name of the disk file and its size. 

  It is possible to get ASCII data out in real-time by using the Disk Recording 

functionality.  

  For users wanting to collect on-line data using a controller; check out the 

System Integrators Guide for information about the data structure. 

 

Basic Principles

  

The Nortek Scour Monitor is 
designed to provide a profile of 
the seabed close to the base of 
a mounting structure. It 
operates as a four beam echo 
sounder; it measures the 
distance to the seabed along 
four downward looking narrow 
acoustic beams with angles 10, 
20, 30 and 45 degrees from 
the vertical. It comes in two 
different versions; one that is 
used for 0.1 to 10 meters 
distance to the bottom (2 MHz) 
and one that is used for 7 to 30 
meters distance (1 MHz). The 
intensity of the acoustic return 
signal along the four beams is 
measured. The bottom 
reflection provides a strong 
peak in signal amplitude, which 
is used to determine the exact 
distance 
 
The basic mechanism causing 
scour is the formation of 
vortices, a result of pile-up of 
water at the upstream side and 
subsequent acceleration of the 
flow around the obstruction, as 
illustrated in the figure below. 

 

 

 

Changes in the seabed level 
are measured in real-time and 
relayed back from the field, 
even during storm periods, 
thus providing invaluable 
information for modelers of 
these changes as they occur. 
The data are corrected for tilt 
and beam angles to provide 
true bottom elevation as a 
function of time. 