background image

3

Suction 

Line

Liquid 

Line

Cap

Schrader 

Valve

Figure 1. Suction & liquid line locations

1. Disconnect all electrical power to the furnace or air handler.

2. Remove system refrigerant per industry standard practices.

3. Disconnect and remove existing evaporator coil.

System Depressurization

1. Remove the cap (Figure 1) from the end of the liquid line.

2. Verify  pressurization  by  depressing  the  Schrader  valve 

on the end of the liquid line. Listen for any escaping gas. 
If there is no pressure, test the coil for leakage.

•  If leakage is found, clearly mark the location of the leak 

and return the coil to the distributor for processing.

•  If no leaks are found, the coil may be installed.

3. Depress the valve to relieve all pressure from the coil.

4. Proceed  to  the  appropriate  lineset  connection  for 

installations with factory installed orifice. See pages 4 - 5.

•  Go to 

Changing the Orifice

 section if your installation 

is with a different orifice.

•  Go to 

Installing a TXV kit

 if your installation is equipped 

with a TXV.

Changing the Orifice

NOte:

 Before proceeding, perform steps 1 - 3 in the System 

Depressurization section and confirm that the restrictor 

orifice size meets the requirements outlined in the outdoor 

unit installation manual. Factory supplied orifice sizes are 

listed in Table 2 (page 4). If the orifice must be replaced, 

follow steps 1 - 5.

 CAUtiON:

to prevent damage to the unit or internal 

components, it is recommended that two 

wrenches be used when loosening or tightening 

nuts. Do not over tighten!

1. Using two wrenches, loosen the nut and distributor body 

as shown in Figure 2. Turn the assembly nut counter-clock-

wise until the orifice body halves are separated.

2. Insert a light-gauge wire hook between the distributor body 

and the restrictor orifice while being careful not to scratch 

either part. Carefully remove the restrictor orifice from the 

distributor body. See Figure 3.

3. Check the actual size of the new orifice. The size is stamped 

on its side. Do not use pin gauges to measure the orifice 

diameter.

4. Insert the new orifice into the distributor body, rounded 

end down. See Figure 4.

 CAUtiON:

to prevent damage to the unit or internal 

components, it is recommended that two 

wrenches be used when loosening or tightening 

nuts. Do not over tighten!

5. Realign the assembly nut on the distributor body and hand 

tighten both components. Mark a line on both bodies and 

then tighten an additional 1/4 turn using two wrenches. The 

movement of the two lines will show how much the nut is 

tightened. If a torque wrench is used, tighten to 10-12 ft. 

lbs. or 14-16 Nm.

6. Proceed to the appropriate lineset connection section. See 

pages 4 - 5.

Installing a TXV Kit

A separate TXV kit and C5 replacement tube kit are required. 

See Tables 3 - 5 (page 5) for proper kit part numbers.

 CAUtiON:

to prevent damage to the unit or internal 

components, it is recommended that two 

wrenches be used when loosening or tightening 

nuts. Do not over tighten!

1. Using two wrenches, loosen the distributor and liquid line 

body halves. Turn the assembly nut counter-clockwise.

2. Discard the removed liquid line.

3. Insert a light-gauge wire hook between the distributor body 

and the restrictor orifice while being careful not to scratch 

either part. Carefully remove the restrictor orifice from the 

distributor body. See Figure 3.

4. Connect the distributor to the outlet side of the valve.

5. Braze the new liquid line from the C5 replacement tube 

kit and liquid line stub & screen (included in TXV kit) with 

dry nitrogen flowing through the joints.

  IMPORTANT: Brazing operation should be performed 

before connecting assembly to the TXV. This protects 

the TXV from heat. Liquid nitrogen prevents internal 

oxidation and scaling from occurring.

 NOte:

 The liquid line is sold separately, see Table 3 (page 

5) for appropriate part number.

6. Connect the liquid line to the inlet side of the valve.

7. Realign the assembly nut on the distributor body and hand 

tighten both components. Mark a line on both bodies and 

then tighten an additional 1/4 turn using two wrenches. The 

movement of the two lines will show how much the nut is 

tightened. If a torque wrench is used, tighten to 10-12 ft. 

lbs. or 14-16 Nm.

8. Using two wrenches, tighten both ends of the valve.

9. Secure the sensing bulb to the suction line using the clamp 

supplied with the kit.

10. Wrap the bulb, clamp, and suction line together with tar 

tape or other insulating material.

impOrtANt NOteS:

•  The sensing bulb must be located flush against the 

suction line for optimum heat transfer.

•  Avoid attaching the sensing bulb to the lowest part of 

the suction line where condensate may accumulate.

•  Do not locate the sensing bulb on vertical sections of 

the lineset.

•  For horizontal lines, the bulb should not be located at 

12 or 6 o’clock position of the suction line. the best 

location is at 4 or 8 o’clock.

Отзывы: