Nortek Aquadopp Profiler Скачать руководство пользователя страница 4

 

Cables 

Our instruments are mostly shipped with RS232 
communication. RS422 is used for cables longer than 
50 m. The wiring is shown in the tables to the right. 
The output can either be in ASCII, in fixed NMEA 
format, or binary format. 
 
Communication baud rate: 300-115200 (user setting). 
Recorder download baud rate: 600/1200 kBaud. 
 
Ensure that the connectors and dummy plugs are 
lubricated with dielectric silicone. We recommend 3M 
silicone spray. Note that the silicone grease included 
in the shipment should be used on the O-rings only.  

 
Analog Input (default for most instruments) 
 

The instrument can read two analog inputs at the 
same time. The input range is 0-5V, corresponding to 
0-65535 counts in the data file. 

 
Power 

 

•  DC Input: 9-16 VDC  
•  Battery DC-input: Nominal voltage: 13.5-18V 
•  Absolute maximum DC input voltage: 18.6 V 
 
 

System Integrator Manual 
 

For more information on how to control a Nortek 
product with a non-PC controller, a System Integrator 
Manual is available on the website. 
 

Troubleshooting 
 

As most problems are caused by simple mistakes, 
please initially check if 
•  you have forgotten to power the system 
•  the connector has fallen out of the computer 
•  you are using the wrong serial port 
 
When all obvious mistakes are ruled out, please turn 
to our 

Troubleshooting Guide

. The FAQ site may 

also be helpful. 

If the problem is still present contact Nortek  
support at [email protected] 

 
 
 
 
 
 
 
 
 
 
 
 

Underwater Connector 

 

Termination 

Pin. No 

Wire Color 

Purpose 

 

Pins 

Description 

black 

RS232 Tx 

Twisted 

Pair 

 

white 

RS232 Rx 

black 

RS232 Gnd 

Twisted 

Pair 

white/purple 

Power Output 

 

black 

Analogue Ch. 2 

Twisted 

Pair 

white/orange 

Analogue Ch. 1 

black 

Power Gnd 

Twisted 

Pair 

white 

Power + 

Screen terminated at pin 1 in underwater connector, unterminated at PC side. 

Ref: N2100-002A 

 

Underwater Connector 

 

Termination 

Pin. No 

Wire Color 

Purpose 

 

Pins 

Description 

black 

RS422 Tx + 

Twisted 

Pair 

 

white 

RS422 Tx - 

black 

RS422 RX - 

Twisted 

Pair 

white/orange 

RS422 RX+ 

black 

Sync. Out 

Twisted 

Pair 

 

white/purple 

Sync. In 

black 

Power Gnd 

Twisted 

Pair 

white 

Power + 

Screen terminated at pin 1 in underwater connector, unterminated at PC side. 

Ref: N2100-108 

 

Underwater Connector 

 

Termination 

Pin. No 

Wire Color 

Purpose 

 

Pins 

Description 

black 

RS232 Tx  

Twisted 

Pair 

 

white 

RS232 Rx 

Black 

Ground 

Twisted 

Pair 

white/purple 

Analogue Z* 

Red wire 

Black 

Analogue X 

Twisted 

Pair 

Green wire 

white/orange 

Analogue Y 

Yellow wire 

black 

Power Gnd 

Twisted 

Pair 

Black wire 

 

white 

Power + 

White wire 

Screen 

Bare 

Ground 

3 bare wires for grounds, connected 

internally to power ground. 

Ref: N2100-004 

 

 

 

Wiring of RS232 with option for analogue outputs. 

*Optional version outputs pressure instead of Z-velocity 

Отзывы: