Specifications –

 

Charger Unit SKL02:

Electrical safety:

EN 60601-1, EN 60950, EN-60335-2-29 

 

EMC standards:

Medical: EN 

60601-1-2 

Emission: EN 

61000-6-3 

Immunity: EN 

61000-6-1 

 

Protection:

Input:  

Fusible resistor 

 

Output: 

Polyswitch fuse against reversed polarity 

 

Temperature range:

÷20 to +40 

°

C

Input:

90 - 264 VAC, 50-60 Hz 

 

Charge voltage:

6 – 12 VDC 

 

Charging method:

Constant current, automatic charge termination. 

 

Charge current:

RC (Rapid Charge):  

2,0A ±200mA 

 

IC (Interim Charge):  

240 mA ±50mA 

 

TC (Trickle Charge): 

55 mA ±25mA 

 

Charging cycle:

RC to IC:
-minus dV (0,5% of Vmax) 

 

-timer: 12 hrs 
 
IC to TC:
-IC timer (2 hrs) reached. 

 

Temp. ranges (NTC):

RC start: Cell temp. < 40

°

C. 

Temperature Cut-Off (TCO) during RC: Cell temp. > 60

°

C. 

 