11
NON ISO 6789-1:2017
NON ISO 6789-1:2017 limit mode allows the user to set upper and lower limits from 0% to 99% of the target
value.
A.
To change the limit target value, use
or
in any measurement mode. The target value set
is shown on the display. Keep either key depressed for fast target value change.
UPPER LIMIT = % OF TARGET
LOWER LIMIT = TARGET - % OF TARGET
The default % OF TARGET is set to 4%. See the SET UP section to change the default value.
Operation
Measurement Mode
Limit Operation
Track
Limits follow signal and are not held.
Dial & Electronic
Limits status is held until ‘ZERO’ is pressed.
Click & Cam
Limit status is held until after the auto reset timer has operated.
Limits are enabled to work in both the Clockwise and Counter clockwise directions.
Status Indication
Torque Signal
Display
LED’s
Serial Port
Under lower limit
LO
Yellow
LO
Within limits
OK
Green
OK
Above upper limit
HI
Red
HI
DIAL &
ELECTRONIC
0000
.
N∙m
TARGET
59.9
UNITS
-
+
ZERO
%