norbar Pro-Test 1500 SERIES 2 Скачать руководство пользователя страница 13

 

 

11 

NON ISO 6789-1:2017 

 

 

 
NON ISO 6789-1:2017 limit mode allows the user to set upper and lower limits from 0% to 99% of the target 
value. 
 

A.

   To change the limit target value, use  

or  

in any measurement mode. The target value set 

is shown on the display. Keep either key depressed for fast target value change. 

 
UPPER LIMIT =  % OF TARGET 
 
LOWER LIMIT = TARGET - % OF TARGET 
 
The default % OF TARGET is set to 4%. See the SET UP section to change the default value. 

 
Operation 

 

Measurement Mode 

Limit Operation 

Track 

Limits follow signal and are not held. 

Dial & Electronic 

Limits status is held until ‘ZERO’ is pressed. 

Click & Cam 

Limit status is held until after the auto reset timer has operated. 

 
Limits are enabled to work in both the Clockwise and Counter clockwise directions. 

 
Status Indication 

 

Torque Signal 

Display 

LED’s 

Serial Port 

Under lower limit 

LO 

Yellow 

LO 

Within limits 

OK 

Green 

OK 

Above upper limit 

HI 

Red 

HI 

 

 

DIAL &

 

ELECTRONIC

 

  0000 

N∙m 

    
TARGET 
      59.9 

UNITS 

 PRINT 

 

ZERO 

 

      % 

A. 

+

 

Содержание Pro-Test 1500 SERIES 2

Страница 1: ...norbar com au NORBAR TORQUE TOOLS INC 36400 Biltmore Place Willoughby Ohio 44094 USA Tel 1 866 667 2279 Email inquiry norbar us NORBAR TORQUE TOOLS PTE LTD 194 Pandan Loop 07 20 Pantech Business Hub...

Страница 2: ......

Страница 3: ...Measurement Mode 7 Fitting Hex Drive Adapter 7 Operating Tool to be Calibrated Tested 7 Set Up 8 Language 8 Limits 8 Units 9 Serial Port 9 Thresholds 9 Modes 9 Limits 10 ISO 6789 1 2017 10 NON ISO 678...

Страница 4: ...1500 series 2 TRANSDUCER DISPLAY Assembly Diagram Part Number Description 29034 sq to 10 mm hex socket 29067 sq to 10 mm hex socket 29120 sq to 10 mm hex socket 29079 sq to 22 mm hex socket 29134 sq...

Страница 5: ...m hex socket 29079 sq to 22 mm hex socket 29134 sq to 22 mm hex socket 29189 sq to 36 mm hex socket 29179 Operators Manual 34299 Calibration certificate Quick reference card s 34302 Serial data lead 3...

Страница 6: ...tolerances The pictorial keys allow direct access to measurement modes Serial Port Connector for data output to PC s or printers PRINT NO PRINT PRINT X PRINT feature for control of serial port output...

Страница 7: ...the mounting bracket part number 62198 not included IMPORTANT THE WHOLE OF THE TRANSDUCER MUST ALWAYS BE IN CONTACT WITH THE SUPPORTING SURFACE Clockwise Only Calibrating Testing Torque tighten to 20...

Страница 8: ...4 Switching On Switch ON Pro Test I in the down position WARNING ALLOW PRO TEST TO EQUALISE TO THE AMBIENT TEMPERATURE HUMIDITY BEFORE SWITCHING ON WIPE OFF ANY MOISTURE BEFORE USE TIP If there is no...

Страница 9: ...TRONIC Bending beam wrenches Dial type wrenches Hold highest signal Reset by pressing CLICK CAM Preset click type wrenches Captures the FIRST PEAK of torque and displays this for the selected AUTO RES...

Страница 10: ...for next menu Press EXIT to end set up LANGUAGE OK EXIT ENGLISH DANSK FRANCAIS NEDERLANDS DEUTSCH SUOMI ITALIANO NORSK ESPANOL SVENSKA PORTUGUES LIMITS OK EXIT X ISO 6789 1 2017 X NON ISO 6789 1 2017...

Страница 11: ...lb gf cm lbf ft in oz lbf in Kgf m X MODES OK EXIT DIAL ELECTRONIC CLICK CAM X SERIAL PORT OK EXIT PARITY OFF DATA STOP BITS 8 2 BAUD RATE 9600 OUTPUT LIMITS YES EVEN ODD OFF SERIAL PORT OK EXIT OUTPU...

Страница 12: ...t value x 100 Target value If the 100 target value is set above 10 N m the deviation must be within 4 For a 100 target value of 10 N m or below the deviation must be within 6 Example For a 100 N m tar...

Страница 13: ...ARGET is set to 4 See the SET UP section to change the default value Operation Measurement Mode Limit Operation Track Limits follow signal and are not held Dial Electronic Limits status is held until...

Страница 14: ...haracter sent before the data Output units YES or NO YES Torque units sent after data SET TO FACTORY DEFAULTS Reset all settings Maximum number of characters per line 24 Transmitted data voltage level...

Страница 15: ...m LO 1 0335 N m NONE Clockwise 1 0335 N m LO 1 0335 N m Anti Clockwise 1 0335 N m LO 1 0335 N m Pin Connections Pin No Function 1 Not Connected 2 Received data to Pro Test 3 Transmitted data from Pro...

Страница 16: ...is functioning at maximum accuracy are available Cleaning Do not use abrasives or solvent based cleaners Disposal Recycling Considerations Component Material Back Panel Stainless Steel Transducer Sha...

Страница 17: ...ture range 5 C to 40 C Storage temperature range 20 C to 70 C Maximum operating humidity 85 Relative Humidity at 30 C Power supply 90 to 264 Volts a c at 50 60 Hz input 9V 300 mA D C output centre pos...

Страница 18: ...16 NOTES...

Отзывы: