norbar Pro-Test 1500 SERIES 2 Скачать руководство пользователя страница 1

 
 
 
 
 
 
 
 
 
 

 

OPERATOR

’S MANUAL 

PROFESSIONAL TORQUE TESTER 

PRO-TEST 60, 400 & 1500 SERIES 2 

FOR USE WITH PRO-TEST

’S FITTED WITH VERSION 

37701.305 SOFTWARE

 

Part Number 34299 | Issue 5 | Original Instructions (English) 

 

NORBAR TORQUE TOOLS LTD 

Wildmere Road, Banbury, 

Oxfordshire, OX16 3JU 

UNITED KINGDOM 

Tel + 44 (0)1295 270333 
Email [email protected]  

 
 

 

NORBAR TORQUE TOOLS 

45

–47 Raglan Avenue, Edwardstown, 

SA 5039 
AUSTRALIA 
Tel + 61 (0)8 8292 9777 
Email [email protected] 

 
 

 

NORBAR TORQUE TOOLS INC 

36400 Biltmore Place, Willoughby, 
Ohio, 44094 
USA 
Tel + 1 866 667 2279 
Email [email protected] 

 
 

 

 

 

 

 

NORBAR TORQUE TOOLS PTE LTD 

194 Pandan Loop 

#07-20 Pantech Business Hub 

SINGAPORE 128383 
Tel + 65 6841 1371 

Email [email protected] 

 
 

 

NORBAR TORQUE TOOLS (SHANGHAI) LTD 

91 Building-7F, No.1122 North Qinzhou Rd, 
Xuhui District, Shanghai 
CHINA 201103 
Tel + 86 21 6145 0368 
Email [email protected] 

 
 

 

NORBAR TORQUE TOOLS INDIA PVT. LTD 

Plot No A-168, Khairne Industrial Area, 
Thane Belapur Road, Mahape, 
Navi Mumbai 

– 400 709 

INDIA 
Tel + 91 22 2778 8480 
Email [email protected] 

 
 

www.norbar.com 

 

 

 

© Norbar Torque Tools Ltd 2017 

Содержание Pro-Test 1500 SERIES 2

Страница 1: ...norbar com au NORBAR TORQUE TOOLS INC 36400 Biltmore Place Willoughby Ohio 44094 USA Tel 1 866 667 2279 Email inquiry norbar us NORBAR TORQUE TOOLS PTE LTD 194 Pandan Loop 07 20 Pantech Business Hub...

Страница 2: ......

Страница 3: ...Measurement Mode 7 Fitting Hex Drive Adapter 7 Operating Tool to be Calibrated Tested 7 Set Up 8 Language 8 Limits 8 Units 9 Serial Port 9 Thresholds 9 Modes 9 Limits 10 ISO 6789 1 2017 10 NON ISO 678...

Страница 4: ...1500 series 2 TRANSDUCER DISPLAY Assembly Diagram Part Number Description 29034 sq to 10 mm hex socket 29067 sq to 10 mm hex socket 29120 sq to 10 mm hex socket 29079 sq to 22 mm hex socket 29134 sq...

Страница 5: ...m hex socket 29079 sq to 22 mm hex socket 29134 sq to 22 mm hex socket 29189 sq to 36 mm hex socket 29179 Operators Manual 34299 Calibration certificate Quick reference card s 34302 Serial data lead 3...

Страница 6: ...tolerances The pictorial keys allow direct access to measurement modes Serial Port Connector for data output to PC s or printers PRINT NO PRINT PRINT X PRINT feature for control of serial port output...

Страница 7: ...the mounting bracket part number 62198 not included IMPORTANT THE WHOLE OF THE TRANSDUCER MUST ALWAYS BE IN CONTACT WITH THE SUPPORTING SURFACE Clockwise Only Calibrating Testing Torque tighten to 20...

Страница 8: ...4 Switching On Switch ON Pro Test I in the down position WARNING ALLOW PRO TEST TO EQUALISE TO THE AMBIENT TEMPERATURE HUMIDITY BEFORE SWITCHING ON WIPE OFF ANY MOISTURE BEFORE USE TIP If there is no...

Страница 9: ...TRONIC Bending beam wrenches Dial type wrenches Hold highest signal Reset by pressing CLICK CAM Preset click type wrenches Captures the FIRST PEAK of torque and displays this for the selected AUTO RES...

Страница 10: ...for next menu Press EXIT to end set up LANGUAGE OK EXIT ENGLISH DANSK FRANCAIS NEDERLANDS DEUTSCH SUOMI ITALIANO NORSK ESPANOL SVENSKA PORTUGUES LIMITS OK EXIT X ISO 6789 1 2017 X NON ISO 6789 1 2017...

Страница 11: ...lb gf cm lbf ft in oz lbf in Kgf m X MODES OK EXIT DIAL ELECTRONIC CLICK CAM X SERIAL PORT OK EXIT PARITY OFF DATA STOP BITS 8 2 BAUD RATE 9600 OUTPUT LIMITS YES EVEN ODD OFF SERIAL PORT OK EXIT OUTPU...

Страница 12: ...t value x 100 Target value If the 100 target value is set above 10 N m the deviation must be within 4 For a 100 target value of 10 N m or below the deviation must be within 6 Example For a 100 N m tar...

Страница 13: ...ARGET is set to 4 See the SET UP section to change the default value Operation Measurement Mode Limit Operation Track Limits follow signal and are not held Dial Electronic Limits status is held until...

Страница 14: ...haracter sent before the data Output units YES or NO YES Torque units sent after data SET TO FACTORY DEFAULTS Reset all settings Maximum number of characters per line 24 Transmitted data voltage level...

Страница 15: ...m LO 1 0335 N m NONE Clockwise 1 0335 N m LO 1 0335 N m Anti Clockwise 1 0335 N m LO 1 0335 N m Pin Connections Pin No Function 1 Not Connected 2 Received data to Pro Test 3 Transmitted data from Pro...

Страница 16: ...is functioning at maximum accuracy are available Cleaning Do not use abrasives or solvent based cleaners Disposal Recycling Considerations Component Material Back Panel Stainless Steel Transducer Sha...

Страница 17: ...ture range 5 C to 40 C Storage temperature range 20 C to 70 C Maximum operating humidity 85 Relative Humidity at 30 C Power supply 90 to 264 Volts a c at 50 60 Hz input 9V 300 mA D C output centre pos...

Страница 18: ...16 NOTES...