N O R A   B O D E   O X Y J E T   G O   O P E R A T I N G   M A N U A L

Important instructions

2.4. Who should not use the OXYJET GO device? Contraindications

This device has been developed to address the first signs of facial ageing. It is only suitable for the treatment  
of healthy skin.

Do not use the device, if you have skin infections, open wounds, rashes, burns, sunburn, inflammation, eczema, 
psoriasis, rosacea, teleangiectasia (couperose or spider veins), active infections like herpes or active acne  
breakouts in the areas to be treated.

Do not use the device, if you have large moles, tattoos, port wine stains, permanent make-up or any skin  
irritation in the areas to be treated.

Do not use the device, if your skin is still recovering from another cosmetic-medical treatment like chemical or 
mechanical peeling, laser resurfacing, fractional laser treatment, dermal filler injections or any other irritating 
skin treatment.

Do not use the device, if you are taking medicines that affect your skin, e. g. cortisone, hormones, vitamin A  
derivatives.

Do not use the device, if you have skin cancer and unclear skin irregularities. If in doubt, consult your doctor 
before using the device.

Do not use the NORA BODE treatment products and the device, if you suffer from allergic reactions towards 
special cosmetic ingredients contained in the products. For more details on the ingredients of the enclosed 
treatment preparations, please refer to the respective outer packaging. If you are not sure if you can use the 
products, please consult your doctor.

2.3. Device-specific warnings

The OXYJET GO device has been developed exclusively for private use and not for professional use in cosmetic 
and medical practice or any other business purpose.

For hygienic reasons the device should only be used by one person.

The OXYJET GO device has been developed for a maximum of 20 minutes of treatment. This time is sufficient  
to treat all areas of your face described in this manual.

To protect against overloading, the OXYJET GO device has an internal timer. After about 18 minutes of  
operation, the device will point you to the approaching end of the treatment time, as the light ring  
changes its colour from blue to red and a warning tone sounds every 20 seconds. Please switch off the  
device in order to avoid overloading.

Disconnect the device from the mains and allow it to cool down for at least 1 hour before putting it back  
into operation.

After about 19 minutes of operation, the light ring begins to flash red and a warning tone sounds every 10 se-
conds. If you ignore both the hazard warning light and the warning sound, the unit automatically switches itself  
off after 20 minutes and remains in ”stand by“ mode. Your device will not turn on for security reasons.

Disconnect the device from the mains and allow it to cool down  
for at least 1 hour before putting it back into operation.

The OXYJET GO device also has an internal temperature switch to  
protect against overheating. If the device becomes too hot due  
to overload or external influences, it switches off automatically.  
The illuminated ring now lights up permanently red and switching  
on the device is prevented. 

Disconnect the device from the mains and allow it to cool  
down for at least 2 hours before putting it back into operation.

20 minutes!

12

13

Содержание OXYJET GO

Страница 1: ...O N E T W O G O G E T T H E G L O W H E G L O W O N E T N A T U R A L L Y B E A U T I F U L S K I N W I T H O U T N E E D L E S G E N T L E E F F E C T I V E A N D S A F E O P E R A T I N G M A N U A L ...

Страница 2: ...ions please do not hesitate to contact us We wish you every success with your OXYJET GO device N O R A B O D E K O S M E T I K OPERATING MANUAL NORA BODE OXYJET GO 1 edition 11 2018 NORA BODE Kosmetik Copying whether in full or in part requires the prior written permission of NORA BODE Kosmetik Printed in Germany MANUFACTURER S ADDRESS NORA BODE Kosmetik Jutta Lothar Bode GbR Glockengasse 8 65329 ...

Страница 3: ... 19 3 1 The OXYJET GO device Functions 15 17 3 2 The NORA BODE treatment preparations 18 19 4 Preparing the OXYJET GO treatment 20 23 4 1 Commissioning of the device 21 4 2 To practise 22 5 The treatment step by step 24 30 5 1 The treatment areas 24 5 2 The preparation of your skin 25 5 3 The infusion pulsation 26 29 5 4 How often you should use the OXYJET GO device 30 6 After the treatment 31 6 1...

Страница 4: ...bert Germany has impres sively confirmed the strong drug penetration accelerating effect of the Bode OXYJET in test series Different from needle injections all cosmetic problem areas can be successfully treated with the OXYJET The whole face especially the delicate skin around the eyes and mouth neck décolleté and body 5 10 professional treatments are necessary to achieve optimal and long lasting ...

Страница 5: ... apply the OXYJET 1 and 2 preparations especially developed for the treatment and speed up their penetration using the OXYJET GO pulsed pressure device In this way you can build up a deposit of active ingredients in your skin which provides an immediate smoothing effect as well as a long term regenerating effect on your skin appearance What you can expect The key ingredient molecular oxygen can he...

Страница 6: ...ice is not washable Never immerse the device in water or other liquids and never rinse it under running water Only use the power adapter provided with the OXYJET GO device Type UES24LCP 120200SPA Do not use the device and or the power adapter if you notice any damages danger of electric shock Do not touch internal components if the unit is damaged to avoid electric shock Always inspect the device ...

Страница 7: ...cific warnings The OXYJET GO device has been developed exclusively for private use and not for professional use in cosmetic and medical practice or any other business purpose For hygienic reasons the device should only be used by one person The OXYJET GO device has been developed for a maximum of 20 minutes of treatment This time is sufficient to treat all areas of your face described in this manu...

Страница 8: ...US your OXYJET GO set includes a suitable storage station this user manual and two preparations that are perfectly tailored to the treatment OXYJET 1 and 2 Before commissioning your OXYJET GO device please note the information in chapter 2 Important instructions before use 3 1 The OXYJET GO device Functions 1 Nozzle 2 On Off Button 3 Pulse Button 4 Flow Button 5 LED light ring 6 Device input socke...

Страница 9: ...the maximum operating time of 20 minutes approaches the LED light ring changes color from blue to red and warning sounds are emitted Please switch off your device now disconnect it from the mains and let it cool down for at least 1 hour Immediately before reaching the forced shutdown maximum operating time of 20 minutes has been reached the light ring flashes red and further warning sounds are emi...

Страница 10: ...nside out Throughout the whole day the skin is supplied with new elasticity firmness and tone Meiritage consists of root extracts of various Asian herbs The active substance complex intensively mois turizes stimulates the collagen synthesis and can provide an evenly pigmented complexion Squalane and Glycosphingolipids fill up the skin s own lipids regenerate the cell membranes and provide a soft a...

Страница 11: ...ower adapter de energized before complete assembly First select the appropriate AC adapter for your country for example the Euro adapter for use in Germany Slide the appropriate AC adapter onto the power adapter from above until it snaps into place Pay attention to the correct and tight fit of the adapter Note To change the AC adapter press PUSH on the power adapter and pull the AC adapter upwards...

Страница 12: ...link Stay on the same skin area for approx 3 pressure surges Now slide on with the nozzle for approx 1 cm and stay on the same skin area again for approx 3 pressure surges Proceed in this way so that you treat a skin area completely Apply gentle pressure to the skin and slowly slide over the skin without completely sealing the nozzle You use the device properly if you feel a gentle clapping on you...

Страница 13: ...asticity fine lines and pigmentation irregularities of your face The following graphic shows you the areas in which you can apply the OXYJET GO treatment Forehead Work towards the hairline from the middle of the forehead Start at the root of the nose and then work upwards line by line Eye area Please note Do not treat directly near the eyes Keep at least 2 cm away from the eye Only treat the area ...

Страница 14: ... some of the product comes into contact with the eye wash out with plenty of water 2 After applying the OXYJET 2 serum tighten the skin area to be treated between the fore and middle fingers and apply the OXYJET GO nozzle with gentle pressure Press the Pulse button and start to work from the inner corner of the eye proceeding step by step to the temple Stay on the same skin area for about 3 pressu...

Страница 15: ...you have treated your entire upper mouth When done press the Pulse button again and the unit will stop pulsation Put your device back in the storage station Now repeat the whole procedure apply the product and pulse it in on the other side of the upper part of the mouth 3 Now distribute 1 2 drops of OXYJET 2 serum below the lower lip and also on the edge of the lip Proceed by treating one half of ...

Страница 16: ... of active ingredients in your skin For an immediate effect if you want to freshen up quickly for a special occasion you can use the device at any time in between 5 4 How often you should use the OXYJET GO device 6 AFTER THE TREATMENT 6 1 The final care of your skin After you have finished the treatment with your OXYJET GO device you should provide your skin with your daily day night care Both NOR...

Страница 17: ... and power adapter with a damp cloth when de energized OXYJET treatment preparations Please contact the authorized OXYJET GO distributor in your country Spare parts like nozzles and power adapter Please contact the authorized OXYJET GO distributor in your country www norabode com otherwise your OXYJET GO device may be damaged Please also note the instructions on the en closed data sheet of the pow...

Страница 18: ...ling damage caused by failure to comply with the instructions in this operating manual accidental or physical damage defects caused by external influences In such cases the User will be held fully liable Adjustment or modi fication of the equipment by anyone other than our authorized representatives will also render this warranty null and void In the case of a valid warranty claim the authorized O...

Страница 19: ...ower adapter is defective The OXYJET GO device is defective The automatic switch off function after 20 minutes of operation has activated The device has switched off automatically because of overheating The DC plug of the power adapter has come loose from the DC socket of the device The wall socket is defective The power adapter is defective The OXYJET GO device is defective The nozzle is clogged ...

Страница 20: ...xposure to electromagnetic fields EMC Errors and changes excepted N O R A B O D E O X Y J E T G O O P E R A T I N G M A N U A L Recycling and technical data 11 RECYCLING This symbol means that neither the device nor the power adapter may be disposed of with normal household waste 2012 19 EU If you want to dispose of the OXYJET GO device please deliver it to the nearest electronic waste collection ...

Страница 21: ...W W N O R A B O D E C O M NORA BODE Kosmetik Glockengasse 8 65329 Hohenstein GERMANY Tel 49 0 6120 6484 Fax 49 0 6120 6329 E Mail info norabode com O N E T W O G O G E T T H E G L O W H E G L O W O N E T ...

Отзывы: