11

NOTE:

 Correct positioning of the device on your finger is critical for accurate measurements. 

While on the finger, do not press the 

GO

2

 

Achieve

 

 against any surface and do not squeeze 

or hold it together. The internal spring provides the correct pressure; additional pressure may 
cause inaccurate readings.
 

GO

2

% SpO

2

Содержание G02 ARCHIEVE

Страница 1: ...User Guide...

Страница 2: ...roduction 1 Contents of Package 1 Symbols 2 Intended Use 4 Installing the Battery 8 Applying the GO2 Achieve to Your Finger 10 Activating the Device 12 Verifying Operation 12 Reading Your Results 13 C...

Страница 3: ...iiENGLISH Troubleshooting 16 Parts and Accessories 19 Specifications 20 Equipment Response Time 23 Testing Summary 24 SpO2 Accuracy Testing 24 Low Perfusion Testing 24 Warranty 25...

Страница 4: ...on about your oxygen saturation the amount of oxygen in your blood and pulse rate at your fingertip The GO2 Achieve is easy to use and needs no routine maintenance except battery replacement This User...

Страница 5: ...No 93 42 EEC for medical devices Type BF Applied Part patient isolation from electrical shock Not for Continuous Monitoring no alarm for SpO2 C LASSIFIE D US C UL UL Mark for Canada and the United St...

Страница 6: ...IP33 Protected against spraying water and against access to hazardous parts with a tool per IEC 60529 SN Serial Number Sensor disconnect the pulse signal is not detected or there is excessive motion...

Страница 7: ...2 5 cm thick The index finger pointer finger is most recommended Warnings Keep the oximeter away from young children Small items such as the battery door battery and lanyard are choking hazards Certai...

Страница 8: ...l not sound if the amount of oxygen in your blood is low or if your pulse rate is too high or too low Do not place the GO2 Achieve in liquid or clean it with agents containing ammonium chloride isopro...

Страница 9: ...ries can leak or explode if used or disposed of improperly Remove the battery if the GO2 Achieve will be stored for more than 30 days Do not use the GO2 Achieve in a combustible environment oxygen enr...

Страница 10: ...and FM radio broadcast towers and TV broadcast towers may affect accuracy Use in emergency vehicles with communication systems may affect accuracy Functional tester cannot be used to assess the accur...

Страница 11: ...s included with each new GO2 Achieve NOTE You may use rechargeable batteries however they may require more frequent replacement 1 Remove the battery door located on the left side of the GO2 Achieve by...

Страница 12: ...Do not force it into place it fits only when positioned correctly When battery is low the battery indicator symbol on the display will flash Remove battery if the device will be stored for more than 3...

Страница 13: ...e opening at the bottom of the device as shown at right until the fingertip touches the built in stop guide The index pointer finger is recommended Have your forearm parallel to the floor when you use...

Страница 14: ...is critical for accurate measurements While on the finger do not press the GO2 Achieve against any surface and do not squeeze or hold it together The internal spring provides the correct pressure add...

Страница 15: ...d Crystal Display appear during the startup sequence Verifying Operation The GO2 Achieve LCD has an integrated backlight that turns on automatically in low light conditions This allows the display to...

Страница 16: ...e rate signal Bars will display after the indicating pulse signal strength the greater the number of bars indicates a greater pulse quality signal strength If you are not getting a pulse rate reading...

Страница 17: ...y indicator symbol shows the battery strength the less the symbol is filled the less battery capacity is available see chart below Replace the battery when this symbol begins to flash Pulse Quality In...

Страница 18: ...on Do not use undiluted bleach or any cleaning solution other than those recommended here as permanent damage could result Dry with a soft cloth or allow to air dry Clean once per week or more frequen...

Страница 19: ...t appear to change you should see a change to the pulse indicator if the device is on the finger Reposition finger or change fingers Remove and replace battery If the problem persists contact Nonin Te...

Страница 20: ...is installed backwards the unit will not function Dead battery Replace battery If the problem persists remove the battery and contact Nonin Technical Service Device may be too cold to operate Allow d...

Страница 21: ...readings Low pulse quality no reading If the indicator is visible and the pulse quality bar graph does not show more than 2 bars try the following 1 Reposition finger 2 Warm finger by rubbing 3 Select...

Страница 22: ...table holder WARNING Certain activities may pose a risk of injury including strangulation if the lanyard should become wrapped around your neck Use the lanyard with caution For more information about...

Страница 23: ...w Perfusion Oxygen Saturation Declared Accuracy Range Arms 70 to 100 SpO2 2 digits Pulse Rate Declared Accuracy Range Arms 20 to 250 BPM 3 digits Low Perfusion Pulse Rate Declared Accuracy Range Arms...

Страница 24: ...1 C as measured during a controlled environment test Humidity Operating 10 to 90 relative humidity non condensing Storage Transportation 10 to 95 relative humidity non condensing Operating Altitude Up...

Страница 25: ...tems This standard is designed to provide reasonable protection against harmful interference in a typical installation However because of the proliferation of radio frequency transmitting equipment an...

Страница 26: ...onse Time 100 95 90 85 80 75 70 65 60 0 0 8 0 16 0 24 0 32 0 40 0 48 0 56 0 64 0 72 0 80 0 SaO2 Reference 4 Beat Average Time in seconds SpO 2 Specific to this example The response of the 4 beat avera...

Страница 27: ...with a laboratory co oximeter The accuracy of the sensors in comparison to the co oximeter samples measured over the SpO2 range of 70 100 Accuracy data is calculated using the root mean squared Arms...

Страница 28: ...vice with a suitable alternative This warranty excludes cost of delivery to and from Nonin Nonin reserves the right to charge a fee for a warranty repair request on any GO2 Achieve found to be within...

Страница 29: ...3700 1st Avenue North Plymouth MN 55441 5443 763 553 9968 877 577 2635 web site www nonin com e mail info nonin com MPS Medical Product Service GmbH Borngasse 20 D 35619 Braunfels Germany 2016 Nonin M...

Страница 30: ...e You ll need this number if you have to contact Nonin with technical service issues or if you have any questions regarding the use or performance of your pulse oximeter Nonin s technical service depa...

Страница 31: ...28 ENGLISH...

Страница 32: ...GO2 1 www go2nonin com 70 C 30 C GO 2 SpO 2 G O 2 SpO 2 2 Warning Keep away from small children this device contains small parts that may pose a choking hazard...

Отзывы: