18
Using the 9847 Pulse Oximeter
2. Release the audible alarm disable button when “
888 888
” is displayed on the %SpO
2
and
pulse rate displays. After the power on self-test, the “audible alarm disabled” indicator will
illuminate steadily. The patient alarms will be set at the default limits. The audible alarms can be
re-enabled by pressing the audible alarm disable button again.
Table 2: Patient Alarm Limit, Breath Beep Pitch, Calendar, and Clock Mode Parameters
Parameter
Appears in
SpO
2
Display
Pulse Rate
♥
Display
Range of Values
Default Value
Alarm Mode
(
dFt
=default limits;
rCL
= recall stored limits;
C02
= only no breath and the sensor
alarms)
ALr
dFt
,
rCL
,
C02
dFt
No Breath Delay (seconds)
nbd
15 to 60 by 5, OFF
20
Oxygen Saturation Low Limit (%)
02L
50 to 95 by 1, OFF
85
Pulse Rate Low Limit (beats/minute)
H
L
30 to 110 by 5, OFF
50
Pulse Rate High Limit (beats/minute)
H
H
75 to 275 by 5, OFF
200
Oxygen Saturation High Limit (%)
02H
a
a.The default setting for “
02H
” is “OFF” (disabled). Any of the other patient alarm limit parameters can be individually
disabled by setting to “OFF”.
80 to 100 by 1, OFF
OFF
Breath Beep Pitch (fixed or variable)
FPt
ON or OFF
OFF (fixed pitch off)
Year
y
00
-
99
09
Month
nn
00
-
12
00
Day
d
01
-
31
00
Hours
h
00
-
23
00
Minutes
nn
00
-
59
00