ThIs guIde Is for refereNce oNLY. refer to the Instructions for use for complete  

information, warnings, cautions and specifications for the product you are using. 

P/N 9477-101-01

8000Q2 

SpO

2

 Ear Clip Sensor

ADDITIONAL GUIDANCE

A.

 Check sensor position and skin at site every 6 to 8 hours.

B.

 If signal quality is low on the oximeter:

    1. Check connection at oximeter.
    2. Check that ear is warm and has good blood flow.
    3. Use the other ear.

C.

 Clean sensor between patients or as needed.

    1. Use soft cloth with 10% bleach/90% water solution ( household bleach < 

10% sodium hypochlorite ).

    2. Do not use ammonium chloride-based cleaning fluids.
    3. Do not overextend the hinge.
    4. Wipe away from connector plug. Do not submerge or immerse sensor or 

connector plug.

    5. Allow to completely dry before next use.