Step 4: Fine Tuning
Press
M90
. This will display Transit time ratio –
R
, Signal Strength-
S
, Signal Quality-
Q
.
For optimum performance you should be with the values:
R: 97% ~ 103%
S: 600 ~ 999
Q: 60 ~ 99
If you do not meet the above values, verify the parameters in Step 2. If everything is
programmed correctly evaluate your application and transducer locations. For more details
refer to the User Manual.
Step 5: Display Flow
Press
M1
to return to the main screen that will display
Total, Rate, Velocity, Signal Strength & Quality.