NOMAD
¤
NexGen Battery
Product Code No. AE650-059
Carefully read these instructions and keep them in a safe place for future reference.
Use the battery only as described in this instruction manual or in the operating instructions
supplied with the
NOMAD
¤
NexGen
nebuliser system
IMPORTANT SAFEGUARDS
- Before using a new battery, charge for at least 24 hours. After initial 24 hour charge,
fully discharge the battery and then recharge it for another 24 hours.
- Subsequent charging time is about 6 hours.
- Battery must be charged and fully discharged at least 5 times in order to condition
the battery to ensure normal battery life. Failure to do this may shorten battery life
and effect the performance of the device.
- At an ambient temperature of 25°C, a fully charged battery will last for about 45
minutes. In the event of long periods of disuse, the battery must be recharged at
least every 3 months to prolong battery life.
- Remove battery from unit, when it is not to be used for extended periods of time.
- Charge the battery only when it is low in order to preserve its storage capacity and
minimise the charging cycles.
- The battery life is equivalent to about 300 charging cycles at an ambient temperature
of 25°C. To ensure that the unit will continue to perform consistently, it is advisable
to replace the battery after this period.
- Do not short-circuit or do not dispose of in fire.
- The case of the battery is not protected against the ingress of liquids. Never wash
1.