USER INTERFACE

This chapter describes how to use the small 

display and buttons inside the display unit to 

perform configuration. The contents of the 

configuration menu is described here too.
The small display inside the case and the large 

display always show the same contents, except 

that the large display may have less than 6 digits 

and then it shows dashes if the content does not 

fit.

The displays may be in the following states:

Normal state – displaying the readings.

Configuration state – changing the settings.

Normal state

After power-up, the unit is in the normal state, 

displaying the input readings according to the 

configuration.

The A3 indicator LED is lit when the external switch 

controlling the FLT or Tare function is active.

The tare function can be used also by pressing the 

*

 key, and the tare value can be reset by pressing 

*

 and 

>

 together (2211KCI only).

Configuration state

Entering

Press the 

*

 and 

^

 keys simultaneously two 

seconds in the normal state to enter the 

configuration state. When entered, the Conf led will 

light.

Navigating

Use the 

^v

 keys to browse the menu. To see a 

configuration value and possibly edit it, push the 

key. Return to the menu with the 

*

 key.

Editing

Integer values

, such as the Dec setting, are 

adjusted with the 

^v

 keys.

Floating point values

, such as scaling and the 

lopass filter, are edited with 

^v>

 keys: select digit 

to edit (blinking) with 

>

 and change it with 

^v

When the decimal point is blinking, it can be 

moved with 

^v

. The first digit can be replaced with 

a minus sign.

Exiting

When all settings are done, exit from the 
configuration menu with 

*

 key. Two options are 

shown: Save to keep the settings made, and Undo, 
to discard all changes. Select 

^v

 Save or Undo 

and push 

>

. (Or push 

*

 to get back to the menu.)

2011KCI configuration menu

Dec

Number of digits after the decimal point. E.g. with value of 1, the reading is displayed with one 

digit after the decimal point.
If the dec setting is negative, the corresponding number of last digits is rounded to zero. 
Example: Dec= 

-

2, the display will step 0, 100, 200...

Lo

Displayed reading at the minimum input signal (2 V). Almost always 0.

Hi

Displayed reading at the maximum input signal (8 V).

Lopass

Damping, will slow down display fluctuations. Set the time constant (63% of the final change) in 

seconds. Fast operation = 0.

9

Содержание 2011KCI

Страница 1: ...2011KCI 2211KCI Weighing display User manual 18 6 2009 Firmware V2 0 ...

Страница 2: ... be scaled freely and independently on each channel These models have been superceded by 2012KCI in spring 2007 A separate Technical manual is available on request covering more details of the KCI displays Table of contents Introduction 2 Ordering 3 Installing 4 Operation 7 User interface 9 Specifications 11 Manufacturer Nokeval Oy Yrittäjäkatu 12 FIN 37100 Nokia Finland Tel 358 3 3424800 Fax 358 ...

Страница 3: ...en with F chassis The number after the letter defines the number of display digits Input card type These large displays use Nokeval 2000 series input cards Custom input cards 2011KCI and 2211KCI have been designed for KCI Power supply options Before summer 2009 the large field displays have been available with three power supply options Name Voltages 24VDC 24 VDC 48VAC 48 VDC or VAC 230VAC 115 or ...

Страница 4: ...ghtmost digit Now the display assembly may be turned away on its hinges Inside the main case The electronics unit The electronics unit may be opened or detached from the case To open it unscrew the four M3 screws that are holding the cover on its corners If the unit has to be detached unscrew the other screws on its top and bottom sides Normally there is no need to open the electronics unit Cautio...

Страница 5: ...used it should be at least 1 AT Connect the Protective Earth wire to the M4 screw using the Faston connector provided a Faston crimping tool is needed For electrical safety the power and signal wires must be fastened using cable ties or equivalent so that if one signal wire is detached it can not touch any of the power supply wires or vice versa 5 0 0 Slot A Slot B Slot C Power Electronics unit Ma...

Страница 6: ...e main case as explained above Use the small display and four buttons to configure the display unit Using the buttons is explained in chapter User interface Via a programming cable Using the programming cable is possible only on models equipped with circuit board 2012 in slot A The programming cable is attaced to a pin header marked POL on the input card in slot A To access it both the main case a...

Страница 7: ... it How many digits are displayed after the decimal point is defined with the Dec setting in the configuration menu The Dec setting can also be used to show fixed zeros which is useful when displaying large readings e g 10000 kg and the last digits would otherwise be nervous The display can be slowed down with a digital 3rd degree lowpass filter The Dead setting can be used to round small readings...

Страница 8: ...nd Hi 1 and Hi 2 correspond to the full scale weighs of the input channels The unit of the scaling can be freely selected kg ton lbs and there is no need for the display to know it Then negative readings are prevented limited to 0 before summing up the readings A tare value is subtracted A new tare value can be stored using an external switch The 3rd order lowpass filter can be used to attenuate f...

Страница 9: ...igating Use the v keys to browse the menu To see a configuration value and possibly edit it push the key Return to the menu with the key Editing Integer values such as the Dec setting are adjusted with the v keys Floating point values such as scaling and the lopass filter are edited with v keys select digit to edit blinking with and change it with v When the decimal point is blinking it can be mov...

Страница 10: ...0 Lo Offset that is added to the display reading May be negative Almost always 0 Hi 1 Displayed reading at the maximum input signal on channel 1 Hi 2 Displayed reading at the maximum input signal on channel 2 When there is a signal on both channels the sum is displayed Lopass Damping will slow down display fluctuations Set the time constant 63 of the final change in seconds Fast operation 0 Dead D...

Страница 11: ...ltage 48 115 or 230 VAC Voltage range 41 265 VAC or VDC Frequency 45 65 Hz Power 25 W Pre fuse 1 AT not required Protection Class 1 Enclosure Dimensions See chapter Installing Protection IP65 Regulations EMC immunity EN 61326 ESD 61000 4 2 4 kV contact criterion B EM field 61000 4 3 10 V m criterion A analog inputs may deflect 0 1 of range Burst 61000 4 4 any port 2 kV criterion B Surge 61000 4 5 ...

Отзывы: