Connecting the safety interlock
36
If either the door switch or the external bus circuit opens (see
), the
inserted modules detect this and disable emission.
Figure 13
Interlock connected to a door switch - Laser SHUTDOWN
Caution:
Do not short-circuit the LEMO interlock connector. Short-circuiting the in-
terlock circumvents safety regulations and NKT Photonics does not take liability
for any injuries or damage caused by doing so.
Caution:
The switch connected to the interlock must be of an approved type. Fur-
ther, the switch must be installed in a manner so that its operation cannot be fixed
in the open state using a tool.
Warning:
If the interlock is bypassed using an interlock defeater, personnel may
be exposed to hazardous laser radiation. To reduce the risk to personnel, the per-
son or group responsible for operation of the equipment must undertake a risk as-
sessment and provide personnel with appropriate personal protective equipment
and safety training.
On
Logic Circuit
External Bus
Connector
Koheras
ACOUSTIK
Bus Defeater
Set
Shutdown Laser
Normally
Open
Accessory
Energized
Interlock
Monitor
Control unit key switch
LEMO
Connector
Door
Switch
BASIK & BOOSTIK LC
Module Control
Circuitry
2
Содержание Koheras ACOUSTIK
Страница 1: ...Item 800 633 01 Koheras ACOUSTIK PRODUCT GUIDE Multi Channel Modular Laser System...
Страница 10: ...10...
Страница 14: ...14...
Страница 16: ...16...
Страница 26: ...System labels 26...
Страница 28: ...28...
Страница 34: ...Operating and storage environment 34...
Страница 42: ...42...
Страница 78: ...Module overview 78...
Страница 80: ...80...
Страница 82: ...82 Figure 44 Mechanical dimensions 483 00 45 00 302 50 4 50 125 00 2...
Страница 92: ...Installing CONTROL 92...
Страница 95: ......
Страница 96: ...1 Koheras ACOUSTIK Product Guide Revision 1 0 09 2021 W 10456...