NIX QC: MANUAL
Device Set-Up
Connecting to Bluetooth
Before attempting to connect to your Nix QC, be sure the device has
been woken up and has a charge (see page 13). On your Android/iOS
smartphone or tablet, ensure Bluetooth is enabled in your settings. Open
the Nix QC app (see page 12 for app details). The app may request to access
your location (this is necessary for Bluetooth connectivity), accept the
request.
Place your Nix QC next to your smartphone or tablet. Press the inactive
device connection icon at the top left of your screen (indicating
that a device is not connected). If prompted, select the device with the
highest signal strength from the list. Typically it takes 15-20 seconds to
connect, but may take up to 60 seconds. You should see an active device
connection icon when connected.
Setting the Delta E and Scan Settings
The default Delta E type is CIE2000 and default scan settings are D50, 2°,
but these can be customized for your needs. Select the slider icon in
the top right to change these settings. Under “Scan Settings” you’ll find
options to change:
• Illuminant & Observer
• Delta E Type
• Delta E Threshold
Remember: Every time the settings are adjusted your new scan logs will be
recorded in a new batch and previous scans will be cleared from the plot. To
access your data, click on the history icon from the main screen.
Using the Reference Tile
You will periodically be required to confirm that the Nix QC is within
advertised specifications (see pages 20-21). The app may request access to
your camera to scan the QR code accompanying the reference tile (used for
the verification process), accept the request.
The app will remind you to verify your Nix QC device by displaying a
warning icon at the top left of the screen. Press the icon to start the
process. Remember to remove any adapters you may have attached to the
Nix QC. Follow the instructions on the app, which will confirm whether or
not the device is calibrated to the factory settings.
If you neglect to perform this step when the warning icon is present, each
scan will be flagged as being out of spec. You can verify your device at
any time from within the settings (slider icon ) by selecting “Device
Verification” under “Device Information.”
See page 17 for instructions on cleaning the reference tile.
Remember: Always close the reference tile container and store in the
included carrying case to prevent dirt or dust buildup which can impact
future measurements.