10
Combination of bearing ring and bearing
The combination of a bearing ring and a thrust needle bearing should be the diagram below indicates,
and apply grease to the bearing ring and the bearing.
Using a different combination (direction) from the diagram below may cause the damage to the parts.
6-4. Memory function
Before storing data, you must set the automatic resetting to manual in advance as per [6-1]-(6)).
Note Note that data cannot be stored in the TR mode.
(1)
How to store data
This product can store measurement data in its memory. Take measurements by following the
procedure described in 6-2.
[1]
To save the displayed torque, press the “UP” key, and the data is saved in the memory from the
memory number “001”.
If any data has been already saved in the memory, the data is saved in the next memory number.
Note Up to 800 data can be saved. When data exceeds 800, data is overwritten from the
memory number 001.
(2)
How to delete memory data individually
[1]
Press the “MEM/OUTPUT” key and “MEM” is displayed at the bottom of the liquid crystal
panel, and the last saved data number and measurement value are displayed in turn.
[2]
Press the “UP” key or “DOWN” key, the memory number is changed, and you can search data.
Note When you do not perform any operation for 6 sec, the “MEM” display disappears, and
the machine goes back to the measurement mode.
[3]
Press the “CLEAR” key while the memory number and measurement value are displayed in
turn, “CLR” is blinking on the liquid crystal panel.
Note When you do not perform any operation while “CLR” is blinking, the measurement
becomes available. (No data is deleted.)
[4]
Press the “CLEAR” key again while “CLR” is blinking, and “- - - -“ is displayed on the liquid
crystal panel, and the selected data is deleted.
(3)
Deletion by specifying the range of the memory data
[1]
Press the “MEM/OUTPUT” key, and “MEM” is displayed at the bottom of the liquid crystal
panel, and the last saved data number and measurement value are displayed in turn.
[2]
Press the “UP” key or “DOWN” key to display the first data of the data that you want to delete.
[3]
Press the “CLEAR” key while the memory number and measurement value are displayed in
turn, and “CLR” is blinking on the liquid crystal panel.
[4]
When you press the “UP” key, the memory number increases. Press the key until the memory
number of the last data in the range you want to delete is displayed.
[5]
When you press “CLEAR” key while the memory number and measurement value are
displayed in turn, “- - - -“ is displayed on the liquid crystal panel, and the data in the selected
range is deleted.
Note When any data in the memory is deleted, the succeeding data moves up to the deleted
memory position.
(4)
Deletion of all memory data
[1]
Press the “CLEAR” key for 4 sec or more, “ALL” is displayed on the liquid crystal panel.
[2]
Pressing the “CLEAR” key a second time while [ALL] is displayed on the liquid crystal panel will make
[CLR] display. Pushing “CLEAR” again will result in [- - - -] being displayed, deleting all data.
Содержание delvo DLT1173A
Страница 2: ......