Fig.9
3 Press the surface of the cuff to make sure that the hook & loop fastens
securely (fi g.9). If the cuff is wrapped tightly or loosely, inaccurate blood
pressure readings may result.
Fig.10
4 If the arm has taper form, it is recommended to put the cuff on “spiral
way”, as shown in the fi gure (fi g. 10).
Fig.11
5 If you are wearing a shirt that might restrict circulation in your upper
arm or you roll your sleeve up over the upper arm, the blood fl ow will be
restricted, preventing accurate measurement (fi g.11).
PROCEDURE FOR THE MEASUREMENT
IMPORTANT!
The device has sensor keys and clicking is a slight touch of a fi nger.
Moisture, dirt and other objects between the fi nger and panel of the device can aff ect the ability to
respond to the touching keys.
1. Insert AIR PLUG into AIR CONNECTOR before starting a blood pressure
measurement.
Take deep breaths and relax. Do not move, chat or strain your arm or arm during
measurement.
2. Press “START/STOP”. Air is exhausted from the cuff (fi g.12).
3. Automatic infl ation starts (fi g.13).
4. Infl ation stops when the pressure, that is, the displayed value, reaches the
designated value.
Noise Interference Detection
Blood pressure value taken while moving cannot be said to be the
correct value because body movement can aff ect blood pressure.
This product analyzes pulse wave and displays «
» when body motion
is detected. Symbol «
» indicates the results might be aff ected by body
movement.
Fig.12
Fig.13