39
The test
does not
start after
applying
the
sample.
Insufficient blood
sample.
Repeat the test using a
new test strip with
larger volume of blood
sample.
Defective test
strip.
Repeat the test with a
new test strip.
Sample applied
after the device is
automatically
turned off.
Repeat the test with
anew test strip. Apply
sample only when
flashing “ ” appears on
the display.
Defective device.
Please contact
customer service.
The
control
solution
testing
result is
out of
range.
Error in performing
the test.
Read instructions
thoroughly and repeat
the test again.
Control solution
vial was poorly
shaken.
Shake the control
solution vigorously and
repeat the test again.
Expired or
contaminated
control solutions.
Check the expiration
date of the control
solution.