20
6. Match Code in Display with Code on Test Strip
vial label being used. If Codes do not match, see
Coding
.
7. Press . Control Symbol
appears in Meter Display.
Note:
If you decide not to perform
a Control Test, press to
remove Control Symbol.
8. Turn Control bottle upside
down. Squeeze one drop of
Control onto a clean tissue. Wipe
off bottle tip.
9. Gently squeeze a drop on a small
piece of unused aluminum foil
or clear plastic wrap. Dispose
after use.
10. With Test Strip still in Meter,
touch edge of Sample Tip to
drop of Control and allow
drop to be drawn into Test Strip.
Remove Test Strip from drop
when Meter beeps. Meter begins
countdown.
LB0010
9 JUNE 2009
LB0010
9 JUNE 2009
Control
Symbol