6.  Match Code in Display with Code on Test Strip 

vial label being used. If Codes do not match, see 

Coding

.

7.  Press       . Control Symbol 

appears in Meter Display.

Note:

  If you decide not to perform 

  a Control Test, press         to 

  remove Control Symbol.

8.  Turn Control bottle upside 

down. Squeeze one drop of 

Control onto a clean tissue. Wipe 

off bottle tip.

9.  Gently squeeze a drop on a small 

piece of unused aluminum foil 

or clear plastic wrap. Dispose 

after use.

10. With Test Strip still in Meter, 

touch edge of Sample Tip to 

drop of Control and allow 

drop to be drawn into Test Strip. 

Remove Test Strip from drop 

when Meter beeps. Meter begins 

countdown.

LB0010 

9 JUNE 2009 

LB0010 

9 JUNE 2009 

Control 

Symbol

Содержание TRUEtrack

Страница 1: ...and the Nipro Diagnostics logo are trademarks of Nipro Diagnostics Inc A4NPD03 Rev 25 Owner s Booklet Manufactured by Fort Lauderdale FL 33309 U S A Scan this code with your smart phone for additional...

Страница 2: ...reference only not intended as a substitute for complete instructions Please read entire Owner s Booklet and product Instructions for Use before testing APPLY BLOOD SAMPLE 1 2 Twosimplesteps INSERT T...

Страница 3: ...s and dedicated customer service If you have questions about using TRUEtrack products visit our web site at www niprodiagnostics com Instructions for use of the System in a professional multi patient...

Страница 4: ...rity of your results within your target range helps slow or stop complications of diabetes Having most blood glucose results within your target range shows how well a treatment plan is working to cont...

Страница 5: ...rson use ONLY DO NOT share your Meter or your Lancing Device with anyone including family members DO NOT use on more than one person ALL parts of your Blood Glucose Monitoring System could carry blood...

Страница 6: ...ere hypotension severe dehydration hyperglycemia with hyperosmolarity with or without ketosis Low blood glucose hypoglycemia symptoms may be trembling sweating intense hunger nervousness weakness and...

Страница 7: ...UEtrack is recommended for testing of human capillary whole blood only TRUEtrack is not recommended for use with venous samples Perform Quality Control Testing before performing a blood test for the f...

Страница 8: ...considered accurate if within 20 of lab results 5 If patient has recently eaten fingerstick results from TRUEtrack can be up to 70 mg dL higher than venous lab results 6 Do not use venous blood with T...

Страница 9: ...ntrol Test 18 21 Testing your Blood How to Obtain a Blood Sample 22 23 How to Test Blood Glucose 24 25 System Out of Range Warning Messages 26 Set Up Time Date 27 28 Memory Morning Averages 14 and 30...

Страница 10: ...Button Select settings view Morning Average values view results in Memory Button Increase numbers in Set Up add Control Symbol move forward by time date when viewing results in Memory Display Shows te...

Страница 11: ...ere Battery Compartment Use one non rechargeable 3V lithium battery CR2032 positive side up Note See Changing Battery for details Meter Label Contains telephone number for assistance Serial Number Lab...

Страница 12: ...e Display Messages 6 Temperature Symbol see Display Messages 7 Drop Symbol Apply blood or Glucose Control 8 Control Symbol see Glucose Control Test 9 Unit of Measure Note Factory Set Cannot be changed...

Страница 13: ...lood or Glucose Control drop to be drawn into Sample Tip Caution Holding the Test Strip Sample Tip to the blood sample too long after the Meter begins testing may cause inaccurate results Note Do not...

Страница 14: ...istance Expiration Dates EXP Write date first opened on vial label Discard vial and unused Test Strips 120 days after written date or date printed next to EXP whichever comes first Control Range Range...

Страница 15: ...ottle if either 3 months after opening or date printed next to EXP on bottle label has passed whichever comes first Control Level 0 1 or 2 Three levels of TRUEcontrol Control are available We recommen...

Страница 16: ...n is pressed see Set Up and Memory The Meter turns off automatically after the Test Strip is removed from the Test Port or after 2 minutes of non use Meter comes with pre set time and date Before usin...

Страница 17: ...s and insert new Code Chip into Meter before testing 1 Start with Meter off Remove and open Code Chip package from box of Test Strips to be used Match Code printed on Code Chip to Code printed on vial...

Страница 18: ...play 4 Match to Code number on vial label If Codes match begin testing Caution If Code shows in Display reinsert Code Chip and then reinsert Test Strip If Code still shows in Display or if Codes do no...

Страница 19: ...Strip is inserted correctly into the Test Port Insert a Strip into the Test Port The Meter is working properly if the full Display appears then the time appears in the upper part of the Display and th...

Страница 20: ...eter damage is suspected Meter was dropped crushed wet etc Note Three levels of TRUEcontrol Glucose Control Solution are available for Quality Control Testing It is important to perform Control Tests...

Страница 21: ...use Test Strips 120 days past written opened date or past date printed next to EXP whichever comes first Discard out of date products and use new products if either date has passed 3 Swirl or invert...

Страница 22: ...tle upside down Squeeze one drop of Control onto a clean tissue Wipe off bottle tip 9 Gently squeeze a drop on a small piece of unused aluminum foil or clear plastic wrap Dispose after use 10 With Tes...

Страница 23: ...ntrol Symbol in the Display Note When using the meter for the first time check the unit of measure in the meter Display after performing a control or blood test Result must display in the correct unit...

Страница 24: ...lancets or lancing device Lancets are for single use only Do not re use From Fingertip 1 Prepare ngertip by washing hands in warm soapy water Rinse well Dry thoroughly 2 Place end of Lancing Device a...

Страница 25: ...e Important Notes Regarding Forearm Testing7 Check with your Doctor or Diabetes Healthcare Professional to see if forearm testing is right for you Results from forearm are not always the same as resul...

Страница 26: ...ay past written date or date printed next to EXP whichever comes first 2 Wash hands and forearm for alternate site testing Rinse well and dry thoroughly 3 Remove Test Strip from vial Recap vial immedi...

Страница 27: ...untdown soon after touching blood drop to Sample Tip discard Test Strip Repeat test with new Test Strip and new blood drop If problem persists see Troubleshooting 9 After countdown result is displayed...

Страница 28: ...pears in Meter Display If blood glucose result is greater than 600 mg dL Hi appears in Meter Display Always repeat test to confirm Low Lo and High Hi results If results still display Lo or Hi call you...

Страница 29: ...er beeps Set Time 2 The hour flashes To change press or on top of Meter to select the hour Like many alarm clocks to set PM scroll through hours until PM appears in Display Note PM does not display if...

Страница 30: ...ber flashes To change press or to select the day Press to set 6 The year flashes To change press or to select the year Press to set Exit Set up To review settings press Make changes as needed per inst...

Страница 31: ...1 Start with Meter off Press and release Display shows 14 day Morning Average value and then 30 day Morning Average value 2 Meter switches between 14 and 30 day Morning Average values for 2 minutes b...

Страница 32: ...mory 2 Press and release to advance to the first blood glucose result in Memory 3 Press and release again to view the latest Control Test result The Control Test result shows the Control Symbol in Dis...

Страница 33: ...low any liquids to enter the Test Port Wipe Meter with a clean lint free cloth dampened with 70 Isopropyl alcohol Let Meter air dry thoroughly before using to test Do not use bleach to clean the Meter...

Страница 34: ...l 120 days after opening or after date printed next to EXP whichever comes first Use of Test Strips past either date may give incorrect results Close vial immediately after removing Test Strip Use Tes...

Страница 35: ...tainer 5 Turn Meter on If Meter will not turn on check that battery was installed properly If not reinsert battery and try again Call for assistance if problem persists Note Replacing battery may affe...

Страница 36: ...pply drop after Meter beeps Repeat with new Test Strip Call for assistance Reason Test Strip inserted upside down or backwards Test Strip not fully inserted Test Strip Error Dead or no battery Battery...

Страница 37: ...rong Test Strip Used Test Strip or Test Strip outside of vial too long Action Wait 10 minutes for Meter and Test Strips to reach room temperature before testing Move Meter and Test Strips to area betw...

Страница 38: ...on Call for assistance Retest with new Strip If error persists call for assistance If you have symptoms such as fatigue excess urination thirst or blurry vision follow your healthcare professional s a...

Страница 39: ...Out of range Lo Result 20 mg dL Action Turn Meter off and on again If error persists call for assistance About 50 tests can be done before battery must be replaced Replace battery Remove Code Chip and...

Страница 40: ...Yellow Springs Instrument YSI Power Supply One 3V lithium battery CR2032 non rechargable Battery Life Approximately 1 100 tests or 1 year Automatic shut off After two minutes of non use Weight 1 66 o...

Страница 41: ...ips Glucose Oxidase Aspergillus sp 2 5 Units Mediators Buffers and Stabilizers TRUEcontrol Glucose Control Contents Volume 3 mL Water 73 D glucose 0 09 0 20 Viscosity Enhancing Agent 25 Inorganic Salt...

Страница 42: ...equential damages so the above limitations or exclusion may not apply 6 This Warranty gives the user specific legal rights and the user may also have other rights which vary state to state Your Nipro...

Отзывы: